Data Specifications; Table 60 Data Specifications - Omni-M-Pa1664; K.3 Data Specifications - Olympus OmniScan MX User Manual

Multitechnology flaw detector
Hide thumbs Also See for OmniScan MX:
Table of Contents

Advertisement

DMTA-20015-01EN [U8778402], Rev. M, February 2017
K.3

Data Specifications

This section details the data specifications of processing, TCG, and storage (see
Table 60 on page 172).
Number of data points
Real-time averaging
Rectification
Filtering
Video filtering
Number of points
A-scan recording
C-scan recording
Maximum file size
172
Appendix K
Table 60 Data specifications — OMNI-M-PA1664
Up to 8000
2, 4, 8, and 16
RF, full wave, half wave+, half wave−
Low-pass (adjusted to probe frequency),
digital filtering (bandwidth, frequency
range)
Smoothing (adjusted to the probe frequency
range)
Programmable TCG
16: One TCG (time-corrected gain) curve
per channel for focal laws
6000 A-scans per second (512-point, 8-bit
A-scan)
I, A, B, up to 10 kHz
Limited to available internal flash memory:
300 MB
Processing
Storage