Table 13 Pinout for the scanner interface LEMO connector (continued)
Pin
I/O
Signal
5
In
DIN3
6
In
DIN4
7
Out
DOUT1/
PaceOut
8
Out
DOUT2
9
In
PhA
axis 1
10
In
PhB
axis 1
11
In
PhB
axis 2
DMTA-20015-01EN [U8778402], Rev. M, February 2017
Description
Digital input 3.
Programmable input. Can be
configured for different
functionalities. By default, this
input is disabled. Refer to the
OmniScan MXU Software User's
Manual ("Configuring the Digital
Input") for information on
programming this input.
To activate the input, you must
use a high-level signal with a
minimum signal length of 50 ms.
Digital input 4.
Programmable input. Can be
configured for different
functionalities. By default, this
input is disabled. Refer to the
OmniScan MXU Software User's
Manual ("Configuring the Digital
Input") for information on
programming this input.
To activate the input, you must
use a high-level signal with a
minimum signal length of 50 ms.
Digital out 1/pace output
Digital output 2
Encoder 1: phase A/clock/up/
down
Encoder 1: phase B/direction/
a
N.U./N.U.
Encoder 2: phase B/direction/
N.U./N.U.
Current
N/A
N/A
±15 mA
±15 mA
N/A
N/A
N/A
Connector References
Level
TTL
TTL
TTL
TTL
TTL
TTL
TTL