General Specifications; Generator And Receiver; Table 62 General Specifications - Omni-M-Ect4; Table 63 Generator And Receiver Specifications - Omni-M-Ect4 - Olympus OmniScan MX User Manual

Multitechnology flaw detector
Hide thumbs Also See for OmniScan MX:
Table of Contents

Advertisement

L.5.1

General Specifications

Connectors
Number of channels
L.5.2

Generator and Receiver

Table 63 Generator and receiver specifications — OMNI-M-ECT4
Number of channels
Number of generators
Supported voltage
Operational frequency
Number of frequencies
Bandwidth
Maximum load
Number of receivers
Operation frequency
Maximum input signal
Maximum balance offset
Gain
Table 62 General specifications — OMNI-M-ECT4
1 19-pin Fischer eddy current probe
connector
1 BNC connector
1 (BNC connector)
4 (19-pin connector)
1 (BNC connector)
4 (19-pin connector)
2 (Gen. 2 used as internal reference)
12 Vp-p
20 Hz to 6 MHz
2 (default); up to four with custom probe
and special chip ID
The different frequencies are
time-multiplexed.
8 Hz to 5 kHz
10 Ω
1 to 4
20 Hz to 6 MHz
1 Vp-p (minimum gain)
266 mVp-p
With analog gain, 28 dB fixed and 40 dB
adjustable
With digital gain, 22 dB to 98 dB
DMTA-20015-01EN [U8778402], Rev. M, February 2017
Channels
Generator
Receiver
OMNI-M-ECT4 Module Specifications