Data

Digitizing frequency
Acquisition rate
Analog/digital resolution 16 bits
Phase rotation
Filtering
Channel processing
Maximum file size
On internal clock
External pace
On 1 axis or 2 axes
Number of alarm outputs 3
Outputs
186
Appendix M
Table 68 Data specifications — OMNI-M-ECA4-32
Data acquisition
40 MHz
1 Hz to 20 kHz (single coil)
The rate can be limited by the instrument's
processing capabilities, or by delays set by
the multiplexed excitation mode.
Data processing
0° to 360° with increments of 0.1°
FIR low-pass (adjustable cutoff frequency)
FIR high-pass (adjustable cutoff frequency)
FIR band-pass (adjustable cutoff frequency)
FIR band-stop (adjustable cutoff frequency)
Median filter (variable from 2 points to
200 points)
Mean filter (variable from 2 points to
200 points)
Mixing
Merging
Interpolation
Limited to available internal flash memory:
180 MB (or 300 MB optional)
Data synchronization
1 Hz to 20 kHz (single coil)
Yes
Yes
Visual, audio, and TTL signal
Data storage
Alarms