Acquisition Specifications; Data Specifications; Table 56 Acquisition Specifications - Omni-M-Pa3232; Table 57 Data Specifications - Omni-M-Pa3232 - Olympus OmniScan MX User Manual

Multitechnology flaw detector
Hide thumbs Also See for OmniScan MX:
Table of Contents

Advertisement

J.2

Acquisition Specifications

This section details the acquisition specifications of frequency, data display, and
synchronization (see Table 56 on page 167).
Table 56 Acquisition specifications — OMNI-M-PA3232
Digitizing frequency
A-scan (acquisition
mode)
Maximum pulsing rate
Depth in material
Refresh rate
Envelope (echo-dynamic
mode)
On internal clock
External pace
On encoder
J.3

Data Specifications

This section details the data specifications of processing, TCG, and storage (see
Table 57 on page 167).
Number of data points
Real-time averaging
Rectification
100 MHz (10 bits)
Up to 6000 A-scans per second (512-point,
8-bit A-scan)
Up to 10 kHz (C-scan)
59.8 meters in steel (longitudinal wave),
10 ms with compression
0.49 meter in steel (longitudinal wave),
81.9 µs without compression
A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz
Yes: Volume-corrected S-scan (30 Hz)
Synchronization
1 Hz to 10 kHz
Yes
On 2 axes: from 1 step to 65536 steps
Table 57 Data specifications — OMNI-M-PA3232
Up to 8000
2, 4, 8, and 16
RF, full wave, half wave+, half wave−
DMTA-20015-01EN [U8778402], Rev. M, February 2017
Frequency
Display
Processing
OMNI-M-PA3232 Module Specifications
167