Data Specifications; Table 72 Data Specifications - Omni-M2-Pann; N.3 Data Specifications - Olympus OmniScan MX User Manual

Multitechnology flaw detector
Hide thumbs Also See for OmniScan MX:
Table of Contents

Advertisement

Table 71 Acquisition specifications — OMNI-M2-PAnn (continued)
External pace
On encoder
N.3

Data Specifications

This section details the data specifications of processing, TCG, and storage (see
Table 72 on page 193).
Number of data points
Real-time averaging
Rectification
Filtering
Video filtering
Number of points
Range
Maximum slope
A-scan recording
C-scan recording
Maximum file size
Yes
On 2 axes: from 1 to 65536 steps
Table 72 Data specifications — OMNI-M2-PAnn
Up to 8192
PA: 2, 4, 8, 16
UT: 2, 4, 8, 16, 32, 64
RF, full wave, half wave+, half wave−
PA channel: 3 low-pass,6 band-pass, and 4 high-
pass filters
UT channel: 3 low-pass, 6 band-pass, and 4 high-
pass filters (3 low-pass filters when configured in
TOFD)
Smoothing (adjusted to the probe frequency
range)
Programmable TCG
32: One TCG (time-corrected gain) curve per focal
law
PA: 40 dB
UT:
40 dB/10 ns
6000 A-scans per second (512-point, 8-bit A-scan)
I, A, B, up to 10 kHz
Limited to available internal flash memory:
300 MB
OMNI-M2-PA1664/16128/32128/32128PR Module Specifications
DMTA-20015-01EN [U8778402], Rev. M, February 2017
Processing
per step of 0.1 dB
100 dB per step of 0.1 dB
Storage
193

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Omniscan mx2

Table of Contents