DMTA-20015-01EN [U8778402], Rev. M, February 2017

Figure N-5 Fastening hole location and specifications

N.1

Acoustic Specifications

This section details the acoustic specifications of the pulser, receiver, and beam
formation, for both UT and PA modes (see Table 69 on page 190 and Table 70 on
page 191).
Table 69 Acoustic specifications — UT channel using UT connector
Voltage
Pulse width
Fall time
Pulse shape
Output impedance
190
Appendix N
28 mm
162 mm
109 mm
Pulser
95 V, 175 V, and 340 V
Adjustable from 30 ns to 1000 ns; resolution
of 2.5 ns
<10 ns
Negative square pulse
<30 Ω
M6, 7.9 mm deep