Table 25 Pinout for the alarm and I/O connector (continued)
Pin
I/O
Signal
7
Out
DOUT4
8
Out
DOUT3
9
In
DIN4/
ExtPace
DMTA-20015-01EN [U8778402], Rev. M, February 2017
Description
Digital output 4
Digital output 3
Digital input 4/external
pace input.
Programmable input. Can
be configured as generic
input 4, or as external pace
input (high level with a
minimum signal length of
50 ms when used as DIN4,
or 21 µs as ExtPace).
Refer to the OmniScan
software user's manuals
("Configuring the Digital
Input") for information on
programming this input.
Current
Level
±15 mA
TTL
±15 mA
TTL
TTL
Connector References
123