Acquisition Specifications; Table 71 Acquisition Specifications - Omni-M2-Pann; N.2 Acquisition Specifications - Olympus OmniScan MX User Manual

Multitechnology flaw detector
Hide thumbs Also See for OmniScan MX:
Table of Contents

Advertisement

DMTA-20015-01EN [U8778402], Rev. M, February 2017
Table 70 Acoustic specifications — PA channel (continued)
Scan type
Aperture
Number of elements
Number of focal laws
Delay range transmission 0 µs to 10 µs in 2.5 ns increments
Delay range reception
N.2

Acquisition Specifications

This section lists the acquisition specifications for frequency, data display, and
synchronization (see Table 71 on page 192).
.
Table 71 Acquisition specifications — OMNI-M2-PAnn
Digitizing frequency
A-scan (acquisition
mode)
Maximum pulsing rate
Depth in material
Refresh rate
Envelope (echo-dynamic
mode)
On internal clock
192
Appendix N
Sectorial and linear
OMNI-M2-PA1664 = 16 elements
OMNI-M2-PA16128 = 16 elements
OMNI-M2-PA32128 = 32 elements
OMNI-M2-PA32128PR = 32 elements
OMNI-M2-PA1664 = 64 elements
OMNI-M2-PA16128 = 128 elements
OMNI-M2-PA32128 = 128 elements
OMNI-M2-PA32128PR = 128 elements
256
0 µs to 6.4 µs in 2.5 ns increments
Frequency
100 MHz
Up to 6000 A-scans per second (512-point,
8-bit A-scan)
Up to 10 kHz (C-scan)
59.8 meters in steel (longitudinal wave),
10 ms with compression
0.49 meter in steel (longitudinal wave),
81.9 µs without compression
Display
A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz
Yes: Volume-corrected S-scan (30 Hz)
Synchronization
1 Hz to 10 kHz