Acquisition Specifications; Data Specifications; Table 47 Acquisition Specifications - Omni-M-Pa16128; Table 48 Data Specifications - Omni-M-Pa16128 - Olympus OmniScan MX User Manual

Multitechnology flaw detector
Hide thumbs Also See for OmniScan MX:
Table of Contents

Advertisement

G.2

Acquisition Specifications

This section lists the acquisition specifications for frequency, data display, and
synchronization (see Table 47 on page 155).
Table 47 Acquisition specifications — OMNI-M-PA16128
Digitizing frequency
A-scan (acquisition
mode)
Maximum pulsing rate
Depth in material
Refresh rate
Envelope (echo-dynamic
mode)
On internal clock
External pace
On encoder
G.3

Data Specifications

This section details the data specifications of processing, TCG, and storage (see
Table 48 on page 155).
Number of data points
Real-time averaging
Rectification
100 MHz (10 bits)
Up to 6000 A-scans per second (512-point,
8-bit A-scan)
Up to 10 kHz (C-scan)
59.8 meters in steel (longitudinal wave),
10 ms with compression
0.49 meter in steel (longitudinal wave),
81.9 µs without compression
A-scan: 60 Hz; S-scan: 20 Hz to 30 Hz
Yes: Volume-corrected S-scan (30 Hz)
Synchronization
1 Hz to 10 kHz
Yes
On 2 axes: from 1 to 65536 steps
Table 48 Data specifications — OMNI-M-PA16128
Up to 8000
2, 4, 8, and 16
RF, full wave, half wave+, half wave−
DMTA-20015-01EN [U8778402], Rev. M, February 2017
Frequency
Display
Processing
OMNI-M-PA16128 Module Specifications
155

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Omniscan mx2

Table of Contents