Acoustic Specifications

This section describes the acoustic specifications for the pulser, receiver, and beam
formation (see Table 34 on page 138).
Table 34 Acoustic specifications — OMNI-M-PA32128
Voltage
Pulse width
Fall time
Pulse shape
Output impedance
Gain range
Input impedance
System bandwidth
Scan type
Aperture
Number of elements
Number of focal laws
Delay range transmission 0 µs to 10 µs in 2.5 ns increments
Delay range reception
138
Appendix C
Pulser
Low: 45 V
High: 90 V
Adjustable from 30 ns to 500 ns; resolution
of 2.5 ns.
<5 ns
Negative square pulse
<30 Ω
Receiver
0 dB to 74 dB in steps of 0.1 dB, maximum
input signal 1.25 Vp-p
50 Ω
0.53 MHz to 21 MHz, ±10 % (−3 dB)
Beam formation
Sectorial and linear
32 elements
128 elements
256
0 µs to 10 µs in 2.5 ns increments