Acoustic Specifications

This section lists the acoustic specifications for the pulser, receiver, and beam
formation (see Table 49 on page 158).
Table 49 Acoustic specifications — OMNI-M-PA16128PR
Voltage
Pulse width
Fall time
Pulse shape
Output impedance
Gain range
Input impedance
158
Appendix H
Pulser
Low: 40 V
High: 80 V
Adjustable from 30 ns to 500 ns; resolution
of 2.5 ns
<10 ns
Negative square pulse
<25 Ω
Receiver
0 dB to 74 dB maximum input signal
1.32 Vp-p
50 Ω (pulse-echo channel)
150 Ω (pitch-catch channel)