DMTA-20015-01EN [U8778402], Rev. M, February 2017
3.
Make sure that the access doors are closed.
4.
Place the cap on the scanner interface connector.
5.
Make sure that the battery compartment door is closed correctly, and that an
acquisition module has been installed on the OmniScan MX2.
Because the acquisition modules are not IP rated, you must install the module and
instrument rubber protectors before cleaning the casing or acquisition module.
Otherwise, liquid could seep into cracks or flow into the housing and damage circuits.
IMPORTANT
Installing rubber protectors on the acquisition module will not render it IP compliant.
6.
Make sure the left side panel is closed.
7.
To restore the instrument's original finish, clean the casing and module with a soft
cloth.
8.
To remove persistent stains, use a damp cloth with a mild, soapy solution. Do not
use abrasive products or powerful solvents, which could damage the finish.
9.
Once the connector protectors are removed, make sure the connectors are dry
before connecting anything to them. If they are wet, dry them with a soft, dry
cloth, or let them air dry.
To clean the casing and acquisition module (OMNI-M2-PA1664, OMNI-M2-PA16128,
OMNI-M2-PA32128 and OMNI-M2-PA32128PR modules only)
1.
Make sure that the instrument is turned off, and that the power cord is
disconnected.
2.
Install a PA connector cap on the PA connector (see Figure 3-1 on page 53).
OR
Connect a EZ Latch probe to the PA connector.
52