Fan Filter Cleaning Procedure; Figure B-2 The Phased Array Module Connector Guide Location - Olympus OmniScan MX User Manual

Multitechnology flaw detector
Hide thumbs Also See for OmniScan MX:
Table of Contents

Advertisement

DMTA-20015-01EN [U8778402], Rev. M, February 2017
7.
Align the connector of the new module on top of the OmniScan connector, and
then gently press down.
8.
Secure the new module with the retaining hooks.
9.
Tighten the two captive screws manually, or with a flathead screwdriver.
10. Put the batteries back in the battery compartment, or plug in the DC power
adaptor if needed.
11. Connect the phased array probe connector as follows:
a) Align the phased array probe connector correctly by matching its guide pins
with the PHASED ARRAY module connector (see Figure B-2 on page 134).

Figure B-2 The phased array module connector guide location

b) Gently press the phased array probe connector into the PHASED ARRAY
module connector until it is fully inserted.
c)
Tighten the retaining screws located on each side of the phased array probe
connector.
B.4

Fan Filter Cleaning Procedure

This section describes the procedure for cleaning the acquisition module fan filter for
models OMNI-M-PA3232, OMNI-M-PA32128, and OMNI-M-PA32128PR.