DMTA-20015-01EN [U8778402], Rev. M, February 2017
Table 54 Data specifications — OMNI-M-PA32128PR (continued)
Filtering
Video filtering
Number of points
A-scan recording
C-scan recording
Maximum file size
164
Appendix I
Low-pass (adjusted to probe frequency),
digital filtering (bandwidth, frequency
range)
Smoothing (adjusted to the probe frequency
range)
Programmable TCG
16: One TCG (time-corrected gain) curve
per channel for focal laws
Storage
6000 A-scans per second (512-point, 8-bit
A-scan)
I, A, B, up to 10 kHz
Limited to available internal flash memory:
OmniScan MX: 180 MB (or 300 MB
optional)
OmniScan MX2: 300 MB