Table 13 Pinout For The Scanner Interface Lemo Connector - Olympus OmniScan MX User Manual

Multitechnology flaw detector
Hide thumbs Also See for OmniScan MX:
Table of Contents

Advertisement

DMTA-20015-01EN [U8778402], Rev. M, February 2017

Table 13 Pinout for the scanner interface LEMO connector

Pin
I/O
1
In
2
Out
3
In
4
In
68
Chapter 6
Signal
AIN
Analog input
+5 V
External power supply
This power is shared with pin 2
of the serial connector (see
"Serial Connector" on page 66).
DIN1
Digital input 1.
Programmable input. Can be
configured for different
functionalities. By default, this
input is disabled. Refer to the
OmniScan MXU Software User's
Manual ("Configuring the Digital
Input") for information on
programming this input.
To activate the input, you must
use a high-level signal with a
minimum signal length of 50 ms.
DIN2
Digital input 2.
Programmable input. Can be
configured for different
functionalities. By default, this
input is disabled. Refer to the
OmniScan MXU Software User's
Manual ("Configuring the Digital
Input") for information on
programming this input.
To activate the input, you must
use a high-level signal with a
minimum signal length of 50 ms.
Description
Current
Level
±2.5 mA
±5 V
(2 kΩ
input
impedan
ce)
500 mA
N/A
N/A
TTL
N/A
TTL