Table 40 Acoustic specifications — OMNI-M-PA16nnM (continued)
System bandwidth
Scan type
Aperture
Number of elements
Number of focal laws
Delay range transmission 0 µs to 10 µs in 2.5 ns increments
Delay range reception
E.2

Acquisition Specifications

This section details the acquisition specifications of frequency, data display, and
synchronization (see Table 41 on page 147).
Table 41 Acquisition specifications — OMNI-M-PA16nnM
Digitizing frequency
Maximum pulsing rate
Refresh rate
Envelope (echo-dynamic
mode)
On internal clock
OMNI-M-PA1616M and OMNI-M-PA1664M Module Specifications
DMTA-20015-01EN [U8778402], Rev. M, February 2017
0.75 MHz to 18 MHz (−3 dB)
Beam formation
Sectorial and linear
16 elements
16 elements for the OMNI-M-PA1616M
64 elements for the OMNI-M-PA1664M
128
0 µs to 10 µs in 2.5 ns increments
Frequency
100 MHz (10 bits)
Up to 10 kHz (C-scan)
Display
A-scan: 60 Hz; S-scan: up to 40 Hz
Yes: Volume-corrected S-scan (up to 40 Hz)
Synchronization
1 Hz to 10 kHz