Figure 8-4 Attaching A Ferrite Clamp Filter To A Cable (Example Shown With The 4Ch/Mux Cable) - Olympus OmniScan MX User Manual

Multitechnology flaw detector
Hide thumbs Also See for OmniScan MX:
Table of Contents

Advertisement

1.
Attach the ferrite clamp filter on the cable of the Olympus array probe
(ultrasound or eddy current, as applicable), in close proximity to the connector for
the OmniScan MX.
2.
Attach the ferrite clamp filter to the cable of the Olympus 19-pin eddy current
probe (4CH/MUX), in close proximity to the connector for the OmniScan MX.
3.
Attach the ferrite clamp filter to the USB cable, in close proximity to the connector
for the OmniScan MX.
4.
Attach the ferrite clamp filter to the Ethernet cable, in close proximity to the
connector (RJ-45) for the OmniScan MX.
5.
Attach the ferrite clamp filter to the scanner interface cable, in close proximity to
the connector (DE-15) for the OmniScan MX.
6.
Attach the ferrite clamp filter to the alarm and I/O cable, in close proximity to the
connector (DE-9) for the OmniScan MX.
Figure 8-4 on page 99 and Figure 8-5 on page 100 show the applicable connection for
various cables on the OmniScan MX, and the locations where the ferrite clamp filters
must be attached.
Figure 8-4 Attaching a ferrite clamp filter to a cable (example shown with the
DMTA-20015-01EN [U8778402], Rev. M, February 2017
4CH/MUX cable)
Basic Operation
99