Acquisition Specifications; Data Specifications; Table 38 Acquisition Specifications - Omni-M-Ut-Nc; Table 39 Data Specifications - Omni-M-Ut-Nc - Olympus OmniScan MX User Manual

Multitechnology flaw detector
Hide thumbs Also See for OmniScan MX:
Table of Contents

Advertisement

D.2

Acquisition Specifications

This section lists the acquisition specifications for frequency, data display, and
synchronization (see Table 38 on page 143).
Table 38 Acquisition specifications — OMNI-M-UT-nC
Digitizing frequency
A-scan (acquisition
mode)
Maximum pulsing rate
Depth in material
Refresh rate
Envelope (echo-dynamic
mode)
On internal clock
External pace
On encoder
D.3

Data Specifications

This section outlines the data specifications for processing, TCG, and storage (see
Table 39 on page 143).
Number of data points
Real-time averaging
Rectification
OMNI-M-UT-2C, OMNI-M-UT-4C, and OMNI-M-UT-8C Module Specifications
100 MHz (10 bits)
Up to 6000 A-scans per second (512-point,
8-bit A-scan)
Up to 10 kHz (C-scan)
59.8 meters in steel (L-wave), 10 ms with
compression
0.49 meter in steel (L-wave), 81.9 µs without
compression
A-scan: 60 Hz
Yes
Synchronization
1 Hz to 10 kHz
Yes
On 2 axes: from 1 to 65536 steps
Table 39 Data specifications — OMNI-M-UT-nC
Up to 8000
2, 4, 8, 16
RF, full wave, half wave+, half wave−