Acoustic Specifications

This section lists the acoustic specifications for the pulser, receiver, and beam
formation (see Table 40 on page 146).
Table 40 Acoustic specifications — OMNI-M-PA16nnM
Voltage
Pulse width
Fall time
Pulse shape
Output impedance
Gain range
Input impedance
146
Appendix E
OMNI-M-PA1664M
Pulser
Low: 40 V
High: 80 V
Adjustable from 30 ns to 500 ns; resolution
of 2.5 ns
<10 ns
Negative square pulse
<25 Ω
Receiver
0 dB to 74 dB, maximum input signal
1.32 Vp-p
75 Ω