DSA8200 Digital Serial Analyzer and Modules Service Manual
during the calibration such that the combined system will have an OC192
(STM-64) or 10.66 Gb Reference Receiver response.
20 GHz: The sampler bandwidth and response is set during the calibration
such that the combined system will have an Optical Bandwidth (- -6 dB)
>20 GHz.
30 GHz: The sampler bandwidth and response is set during the calibration
such that the combined system will have an Optical Bandwidth (- -6 dB)
>28 GHz.
The 80C04 can be optionally configured with clock recovery (Opt. CR-1) that
supports 9.953 Gb/s telecom standards.
The 80C05 module is designed to test long wavelength (1520- -1580 nm) signals.
This module is intended to be used as a test and measurement tool for high
bandwidth telecommunications with its high optical bandwidth >40 GHz.
OC192: There is one Reference Receiver setup selectable for 9.95328 Gb/s
SONET/SDH standard.
There is no clock recovery option available.
The 80C06 module is designed to test long wavelength (1520- -1580 nm) signals.
This module is intended to be used as a test and measurement tool for high
bandwidth telecommunications with its high optical bandwidth >55 GHz.
55 GHz: There is only a single bandwidth selection available, no Reference
Receiver setups selectable.
There is no clock recovery option available.
The 80C07 module is designed to test both long and short wavelength
(700- -1650 nm) signals. This module is intended to be used as a test and
measurement tool for high bandwidth telecommunications with its high optical
bandwidth >2.3 GHz.
Filtered rates are OC-3, OC-12; unfiltered rate is OC-48.
There is clock recovery option available (155/622/2488 Multi-rate)
The 80C07B module is designed to test both long and short wavelength
(700- -1650 nm) signals. This module is intended to be used as a test and
measurement tool for high bandwidth telecommunications with its high optical
bandwidth >2.3 GHz.
Theory of Operation