Installation
Install the Sampling Modules
Check Your Sampling
Module Manual(s)
Large-module compartments (2)
Small-module compartments (4)
Connect ESD wrist strap here
1- 2
operating requirements, see the DSA8200, TDS8200, TDS8000, TDS8000B,
CSA8200, CSA8000, and CSA8000B Rackmount Installation Kit Instructions.
WARNING. Do not install or remove any sampling modules while the instrument
is powered on. Electrical shock may occur. Always power the instrument down
before attempting to remove or insert any sampling module to avoid potential
injury from shock.
CAUTION. Sampling modules are inherently vulnerable to static damage. Always
observe static-safe procedures and cautions as outlined in your sampling module
user manual.
Read the sampling-module user manual for instructions on how to install your
sampling modules.
NOTE. After first installing a sampling module(s) or after moving a sampling
module from one compartment to another, you should run compensation from the
Utilities menu to ensure the instrument meets its accuracy specifications. You
must run a compensation (accessed from the Utilities menu) whenever the
extender configuration is changed from that present at the last compensation. In
short, if you install or remove an 80E00 extender, run a compensation. If you
exchange an extender for one of a different length, run a compensation.
Figure 1- -1 shows compartments for both large and small sampling modules,
along with the plug-in connector for the ESD wrist strap that you must use to
install and remove these modules.
Figure 1- 1: Compartments for sampling modules
DSA8200 Digital Serial Analyzer and Modules Service Manual