To remove the RIC-07 (Logic input cable) and the RIC-08 (Alligator clip cable), or the RIC-09 (IC clip
cable) after measurement, insert a screwdriver into the connection and remove it by loosening the
indentation.
•
RIC-10 (Probe set) set contents (optional)
• RIC-07 (Logic input cable)
• RIC-08 (Alligator clip cable)
• RIC-09 (IC clip cable)
• Logic Seal
To avoid break-downs or short-circuiting accidents, please make sure to abide by the items written below.
•
Maximum input voltage
• In case the input voltage exceeds the specifications, the circuit at the input part will break down so even if the input
voltage exceeds the specifications only for an instant, please don't input.
• Please take care that no static electricity is applied on the logic/pulse signal input terminal.
<Input terminal Each channel / Common terminal interval>
Maximum input voltage : 26.4Vmax
The input of Logic/Pulse module has been pulled up 10KΩ, however, the module may be affected by the external noise
when the input is opened.
In such cases, please take the following measures.
•
Set the voltage input without having to open the input.
•
Set the filter to ON (approx. 50 Hz) in the input settings.
CHAPTER 2 Checks and Preparation
2-35
Need help?
Do you have a question about the gl7000 and is the answer not in the manual?