R&S EB200 Manual page 93

Miniport receiver
Table of Contents

Advertisement

EB200 Manual
4.5.4 CALCulate subsystem
CALCulate
.
:IFPan
.
.
:AVERage
.
.
.
:TIME <numeric_value>
Setting of the averaging time for the SCALar averaging procedure
Parameter:
<numeric_value>
Remark:
For MAIN CPU versions > 1.18 this command not only has effect on the averaging time of the IF-
panorama but also on the global measuring time. Thus the original time grading (0, 100, 500 ms) is
discarded (see "MEASure subsystem" on page 40").
For reasons of compatibility with versions ≤ 1.18, entering the averaging time = 0 means the same as
the command CALCulate:IFPan:AVERage:TYPE OFF.
*RST state:
0
Example
CALCulate:IFPan:AVERage:TIME 0.1
.
.
.
:TIME?
Query of the set averaging time for the SCALar averaging procedure
Parameters:
none
Result:
Averaging time in seconds
Remark:
For reasons of compatibility, for MAIN CPU versions > 2.31 this command reports the global measuring
time unless CALCulate:IFPan:AVERage:TYPE? is set to OFF.
Example:
CALCulate:IFPan:AVERage:TIME? -> 0.1
4052.2000.02
averaging time in steps of 0, 100, 500 ms
4.19
CALCulate subsystem
E-7

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the EB200 and is the answer not in the manual?

Table of Contents

Save PDF