:FEED? <trace_name>
Query of data block connected with the trace
Error message:
If the trace name is unknown, an error -141, "Invalid character data" will be generated.
Parameters:
<trace_name>
Result:
Name of the block coupled to the trace.
The following is valid for MTRACE: "SENS"
The following is valid for ITRACE: "FREQ"
The following is valid for IFPAN: "SENS"
Example:
TRACe:FEED? MTRACE -> "SENS"
.
.
:CONTrol <trace_name>, ALWays|SQUelch|NEVer
Control of trace loading
Error message:
If trace name is unknown, an error -141, "Invalid character data" will be generated.
Note:
For IFPAN Trace, only ALWays or NEVer can be selected.
Parameters:
<trace-name>
ALWays
SQUelch
NEVer
*RST state:
NEVer
Example:
TRACe:FEED:CONTrol MTRACE, ALWays
.
.
:CONTrol? <trace_name>
Query of trace loading
Error message:
If the trace name is unknown, an error -141, "Invalid character data" will be generated.
Parameters:
<trace_name>
Result:
ALW, SQU, NEV
Example:
TRACe:FEED:CONTrol? MTRACE -> ALW
4052.2000.02
see TRACe[:DATA]?
see TRACe[:DATA]?
all data are stored
data are first stored, if the signal has exceeded the squelch
threshold defined in the OUTPut:SQUelch subsystem
do not store any data in the trace.
see TRACe[:DATA]?
4.115
TRACe subsystem
E-7