Measuring Time - R&S EB200 Manual

Miniport receiver
Table of Contents

Advertisement

EB200 Manual
.
:TIME <numeric_value>|MINimum|MAXimum|DEFault
Setting of the measuring time for all measuring functions
The set measuring time also has an impact on the averaging time of the IF-panorama data.
Note:
As from MAIN CPU version 2.31, the version of the DSP software (IF DSP) indicates what kind of DDC
is fitted in the IF section. When the displayed version is smaller than V03.00, then the IF section
contains a DDC1. For versions from V03.00 onwards the IF section holds a DDC2. The equipment with
DDC2 is a prerequisite for future software options such as EB200CM and the configuration of a

measuring time.

Note:
Selecting a measuring time which really makes sense is the user's responsibility. Too short a
measuring time will give rise to incorrect results.
Parameters:
<numeric_value>
MINimum|MAXimum
DEFault
*RST state:
DEFault
Example:
MEASure:TIME 200 ms
.
:TIME? [MINimum|MAXimum]
Query of the set measuring time
Parameters:
none
MINimum|MAXimum
Result:
Measuring time in seconds; with the default measuring time being set DEF will be output.
Example:
MEASure:TIME? -> 0.2
4052.2000.02
measuring time in seconds
shortest/longest measuring time
preset measuring times
query of the current measuring time
query of the shortest/longest measuring time
4.41
MEASure subsystem
E-7

Advertisement

Table of Contents
loading

Table of Contents