Responses To Queries - R&S EB200 Manual

Miniport receiver
Table of Contents

Advertisement

EB200 Manual

4.4.4 Responses to queries

A query is defined for each setting command unless explicitly specified otherwise. It is formed by adding
a question mark to the associated setting command. According to SCPI, the responses to queries are
partly subject to stricter rules than in standard IEEE 488.2.
1 The requested parameter is transmitted without header.
Example:
SENSe:FREQuency:MODe
2. Maximum values, minimum values and all further quantities, which are requested via a special text
parameter are returned as numerical values.
Example:
FREQuency? MAX
3. Numerical values are output without a unit. Physical quantities are referred to the basic units.
Example:
FREQuency?
4. Truth values <Boolean values> are returned as 0 (for OFF) and 1 (for ON).
Example:
OUTPut:FILTer:STATe?
5. Text (character data) is returned in a short form (see also "Parameters" on page 9).
Example:
SENSe:FREQuency:MODe?
4052.2000.02
Response: 100000000 for 100 MHz
4.8
Remote control
Response: SWE
Response: 3000000000
Response: 1
Response: SWE
E-7