4.5.17 TEST subsystem

The self-test can be run with two different test routines. The basic test runs continuously in the
background and tests the test points inside the module. Based on this test, a "short test" or a "long test"
can be triggered. In the short test, a line spectrum is fed in at the antenna input and the receiver is set
to the line frequency nearest to the receive frequency. The complete receive path from the antenna
input of the tuner to the level evaluation is then measured and evaluated. In the long test, each line
frequency of the test spectrum is set and measured.
TEST? SHORt|LONG, REPort|QUIet
Triggering the "short test" or "long test"
Parameters:
SHORt|LONG
REPort|QUIet
Note:
When the test was triggered with REPort plain text error messages will be stored in the error queue.
They can be queried with SYStem:ERRor? .
Result:
0
≠ 0
Example:
TEST? LONG, QUIET -> 1
4052.2000.02
carry out short test | long test
error messages in plain text are generated | not generated
no error detected
error detected :
4.109
TEST subsystem