EB200 Manual

5.2.2 Error messages

Error message
Backup LI-Battery low
DC Converter defective
DSP IF Section defective
Front Panel defective
Frontend defective
Frontend1 defective
Frontend2 defective
IF Panorama defective
IF Prefilter narrow defective
IF Prefilter wide defective
IF Section defective
Preselector 20..80 MHz defective
Preselector 80..200 MHz defective
Preselector 200..650 MHz defective
Preselector 650..1500 MHz defective
Preselector 1500..3000 MHz defective
Preselector defective
Remote Interface defective
Testgenerator defective
4052.2000.02
Maintenance and Troubleshooting
Cause
The lithium backup battery for storing parameters in
the CMOS RAM is dead.
The DC/DC converter indicates an error.
The IF section (EB200Z1) indicates an error.
The front panel control unit (EB200F1) indicates an
error.
The standard frontend indicates an error.
The standard frontend indicates an error.
The standard frontend indicates an error.
The IF panorama module (EB200SU) indicates an
error. The EB200SU is optional.
The IF section (EB200Z1) indicates an error.
The IF section (EB200Z1) indicates an error.
The IF section (EB200Z1) indicates an error.
The preselection (EB200V1) indicates an error.
The preselection (EB200V1) indicates an error.
The preselection (EB200V1) indicates an error.
The preselection (EB200V1) indicates an error.
The preselection (EB200V1) indicates an error.
The preselection (EB200V1) indicates an error.
The remote control interface (eg EB200R2) indicates
an error.
The preselection (EB200V1) indicates an error.
5.3
Module
LI-battery
A11
A8
A3
A7
A7
A7
A9
A8
A8
A8
A6
A6
A6
A6
A6
A6
A12
A6
E-7.1