Resetting Values Of The Status Reporting System - R&S EB200 Manual

Miniport receiver
Table of Contents

Advertisement

EB200 Manual

4.7.5 Resetting values of the status reporting system

Table 4-12 comprises the different commands and events causing the status reporting system to be
reset. None of the commands, except for *RST, influences the functional device settings. In particular,
DCL does not change the device settings.
Table 4-12 Resetting device functions
Event
Effect
Clear STB, ESR
Clear SRE, ESE
Clear PPE
Clear EVENTt sections of
the registers
Clear ENABle section of all
OPERation and
QUEStionable registers,
Fill ENABle sections of all
other registers with "1".
Fill PTRansition sections
with "1" ,
Clear NTRansition sections
Clear error queue
Clear output buffer
Clear command processing
and input buffer
1) The first command in a line, ie immediately following a <PROGRAM MESSAGE TERMINATOR>
clears the output buffer.
4052.2000.02
Switching on supply
voltage
DCL, SDC
(Device Clear,
Selected Device
Clear)
yes
yes
yes
yes
yes
yes
yes
yes
yes
*RST
yes
1)
yes
4.142
Remote control
STATus:PRESet
*CLS
yes
yes
yes
yes
yes
1)
1)
yes
E-7

Advertisement

Table of Contents
loading

Table of Contents