C H AP TE R 10 - USI N G A D VA N CE D F EA TU R ES
4.
Within 5 seconds, your pump will respond with a number of confirmation beeps/vibration pulses
equal to the number of times you pressed the Audio Bolus button.
DO NOT press any of the function buttons until the confirmation beep sequence is complete. Once
the confirmation beep sequence is complete, you can press any function button other than the Audio
Bolus button or Contrast button to cancel.
5.
Within 5 seconds, your pump will beep/vibrate twice to "ask" you to
confirm that you wish to activate delivery and "Confirm" is displayed on the
Audio Bolus screen.
NOTE: You can press any function button other than the Audio Bolus button or
Contrast button to cancel.
6.
Within 5 seconds, press the button again to activate delivery. Your pump
will beep/vibrate twice to confirm your delivery command. The DELIVERING
bolus screen is displayed and your pump will beep/vibrate once to signal the
start of delivery and once to signal end of delivery (if you turned on Normal
Bolus Sounds in Setup).
NOTE: You can press any function button other than the Audio Bolus button or
Contrast button to cancel.
If you cancel a bolus delivery after you have activated it, the screen at right
will be displayed. See Chapter 11 in Section I.
NOTE: If during a bolus delivery your low cartridge level is reached, your pump will not display the
warning until after the bolus is completed. So you could possibly have less insulin available than your
low cartridge setting after the bolus is delivered.
The maximum number of Audio Bolus button presses is 20. Therefore, if you have set the step size to
0.1U, the maximum audio bolus amount is 2U. If you have set the step size at 0.5U, the maximum
audio bolus amount is 10U and if your step size is 1.0U, the maximum audio bolus amount is 20U.
With a 5.0U step size, the maximum audio bolus amount cannot be greater than 35U, which is the
maximum amount for any type of bolus. You will be alerted on the display if you try to deliver any
bolus amount that exceeds 35U.
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 62
62
Dec 2014
Audio Bolus
Confirm
U
3.00
Press any other
key to CANCEL
DELIVERING
U
3.00
Press any
function button
to cancel.
Warning
Bolus
delivery
canceled by
user button
press.
Delivered:
2.00U of 3.00U
Confirm
A170 of A312
11/14/14 2:52 PM