Dexcom g4 platinum Quick Start Manual page 189

Sensor and transmitter
Hide thumbs Also See for g4 platinum:
Table of Contents

Advertisement

ANIMAS® VIBE™ SYSTEM
C H AP TE R 1 0 - CG M SA F ET Y S Y ST EM AN D A L ER T S
CGM Warning: Enter BG
Cause
Effect
Message
Action
Beeps/Vib
CGM Warning: Always use Fingerstick BG for Treatment Decisions
and CGM Calibration
Cause
Effect
Message
Action
Beeps/Vib
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 161
The BG value entered has not been accepted for
startup calibration or calibration update.
Insulin deliveries continue. CGM continues but with a
possible CGM data gap.
Displayed each time pump is awakened until
confirmed, and action is taken.
to confirm. Take another fingerstick BG
Press
test and enter value in pump. You will continue to
be reminded to enter a valid BG value until the BG
value is accepted for recalibration. DO NOT enter a
fingerstick BG value if
CGM Trend or Data screen.
User selected (under CGM Setup menu options),
every 3 minutes up to 3 times or until confirmed.
No progression.
Appears at start of every CGM session.
Insulin deliveries continue. CGM continues.
Displayed each time pump is awakened until
confirmed.
Press
to confirm. Never use a CGM reading to make
treatment decisions or to calibrate the CGM. Take
fingerstick BG test before adjusting insulin dose,
eating, exercising, or making any other treatment
decisions. Using CGM readings to make treatment
decisions can result in under delivery or over delivery
of insulin. Using CGM readings to calibrate the CGM
can result in inaccurate CGM readings.
User selected (under CGM Setup menu options), one
time. No progression.
161
or
appear on the
Dec 2014
CGM Warning
Enter BG
Confirm
CGM Warning
Always use
Fingerstick
BG for
Treatment
Decisions
and CGM
Calibration
Confirm
A271 of A312
11/14/14 2:53 PM

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the g4 platinum and is the answer not in the manual?

Questions and answers

Table of Contents