ANIMAS® VIBE™ SYSTEM
C H AP TE R 5 - C GM C AL I B RA TI O N
2-Hour CGM Startup Period Calibration
CGM Warning
Enter 2
Startup BGs
Confirm
CGM Menu
BG Cal.
Trend Graph
Start/Stop
History
Setup
Main Menu
BG Calibrate
120
mg/dL
CALIBRATE
CANCEL
3.
Repeat step 2 to enter the second, new BG value. The second BG value must be from a new
fingerstick test from a BG meter. DO NOT re-enter the first BG value again as this may impact the
accuracy of your Sensor glucose readings.
NOTE: Suspending insulin delivery during the 2-hour CGM Startup period will not affect the initial
calibration sequence, and the CGM session will remain active.
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 138
When the 2-hour CGM startup period ends, you will be prompted to enter 2
separate fingerstick BG values in your pump. Press
1.
Scroll to "BG Cal." on the CGM Menu screen and press
2.
The "BG value" field will be highlighted and flashing on the BG Calibrate
screen. Use the
/
buttons to enter the first of 2 fingerstick BG values.
Press
. "CANCEL" will be highlighted. Scroll to "CALIBRATE" and press
If you need to cancel the BG value, scroll to "CANCEL" and press
case, you will return to the CGM Menu.
NOTE: The default value on the BG Calibrate screen is 120 mg/dL the first
time you calibrate.
138