ANIMAS® VIBE™ SYSTEM
C H AP TE R 3 - GE T TI N G Y O U R PU M P R EA D Y
6.
Insert your filled cartridge.
7.
Secure cartridge compartment cap to pump by
turning in a clockwise motion until snug but DO NOT over tighten.
m
WARNING: NEVER start any part of the Prime/Rewind sequence on your pump while the
infusion set is connected to your body. Always disconnect before you begin the sequence.
The Prime/Rewind sequence includes steps for rewinding the pump motor, loading an insulin
cartridge, and priming the infusion set tubing. Failure to disconnect your infusion set from your
body before performing these steps can result in over delivery of insulin.
ezPrime
Rewind
Load Cart
Prime
Fill Cannula
Main Menu
8.
On the REWIND COMPLETE screen, "Continue"
is highlighted. Press
piston rod with the cartridge. The LOAD CARTRIDGE
ACTIVE screen is displayed, followed by the PRIME
screen. Your pump will beep once to let you know
the cartridge is aligned with the piston rod.
PRIME
Be sure set is
disconnected
from your body.
Insulin: 200U
Continue
Cancel
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 21
NOTE:
• If screen display has timed out while loading your cartridge, select "Prime/
Rewind" from the MAIN MENU and highlight "Load Cart" from the ezPrime
menu. Press
to display the REWIND COMPLETE screen. Continue with
Step 8.
• You can only highlight "Load Cart" from the ezPrime menu after the rewind
action is complete. This is true for all ezPrime menu options: you can only
proceed to the next action once the previous action is complete.
. Your pump will align the
9.
On the PRIME screen, "Continue" is highlighted. Press
DELIVER PRIME screen.
21
REWIND
COMPLETE
Disconnect set.
Load U100
cartridge.
Attach cap.
Continue
Cancel
Dec 2014
LOAD
CARTRIDGE
ACTIVE
Please wait.
----
----
----
>>
>>
>>
>>
to display the
A129 of A312
11/14/14 2:52 PM
Need help?