ANIMAS® VIBE™ SYSTEM
B EF O RE YO U BE G IN
Guidelines Involving the Animas
m
The following guidelines cover the pump, Transmitter/Sensor, and infusion sets, and apply to the
patient and/or HCP administering the procedure.
Medical Procedure
• CARDIAC CATHETERIZATION
• CT SCANS
• ELECTRO-CAUTERY
• MRI
• NUCLEAR STRESS TEST
(myocardial perfusion imaging)
• PACEMAKER/AICD Reprogramming
• THERAPEUTIC RADIATION/ONCOLOGY
(cancer treatment radiation)
• X-RAY, DENTAL
• X-RAY, MAMMOGRAM
• COLONOSCOPY
• EKG
• LASER SURGERY
• ULTRASOUND
• X-RAY, BODY FLUOROSCOPY
(chest, neck, abdomen, torso, etc.)
• X-RAY, BONE DENSITOMETRY
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 9
Vibe
System and Medical Procedures
®
™
Guidelines
• DO NOT bring pump or Transmitter/Sensor into
the same room where the procedure is being
performed.
• Teflon/plastic infusion set can remain in. Other
types of infusion sets must be removed.
• Pump, infusion set and Transmitter/Sensor can
remain on/in.
• DO NOT bring pump or Transmitter/Sensor into
the same room where the procedure is being
performed, unless a lead apron that completely
covers pump is worn during the procedure. The
person administering the procedure (if they are
wearing a pump) must proceed to a designated
safety area during the procedure.
• Teflon/plastic infusion set can remain in. Other
types of infusion sets must be removed.
• DO NOT bring pump or Transmitter/Sensor into
the same room where the procedure is being
performed, unless a lead apron that completely
covers pump and Transmitter is worn during the
procedure. The person administering the procedure
(if they are wearing a pump or Transmitter) must
proceed to a designated safety area during the
procedure.
• Teflon/plastic infusion set can remain in. Other
types of infusion sets must be removed.
IX
Dec 2014