Dexcom g4 platinum Quick Start Manual page 195

Sensor and transmitter
Hide thumbs Also See for g4 platinum:
Table of Contents

Advertisement

ANIMAS® VIBE™ SYSTEM
C H AP TE R 1 0 - CG M SA F ET Y S Y ST EM AN D A L ER T S
CGM Warning: CGM Session Stopped
Cause
Effect
Message
Action
Beeps/Vib
CGM Warning: CGM Sensor Failure, Insulin Delivery Continues
Cause
Effect
Message
Action
Beeps/Vib
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 167
CGM session stopped due to a problem with the
Sensor or Transmitter.
Insulin deliveries continue. CGM session stops.
Displayed each time pump is awakened until
confirmed.
Press
to confirm. Replace Sensor and start new
CGM session.
User selected (under CGM Setup menu options),
every 3 minutes up to 3 times or until confirmed.
No progression.
Sensor error indicating that Sensor is not functioning
properly.
Insulin deliveries continue. CGM session ends.
Displayed each time pump is awakened until
confirmed.
Press
to confirm. Contact Customer Service for
assistance.
User selected (under CGM Setup menu options),
every 3 minutes up to 3 times or until confirmed.
No progression.
167
Dec 2014
CGM Warning
CGM Session
Stopped
Confirm
CGM Warning
CGM Sensor
Failure,
Insulin
Delivery
Continues
Confirm
A277 of A312
11/14/14 2:53 PM

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the g4 platinum and is the answer not in the manual?

Questions and answers

Table of Contents