ANIMAS® VIBE™ SYSTEM
C H AP TE R 1 0 - CG M SA F ET Y S Y ST EM AN D A L ER T S
CGM Warning: CGM Failure, Insulin Delivery Continues, Call
Service
Cause
Effect
Message
Action
Beeps/Vib
CGM Warning: Replace CGM Transmitter, Low Battery
Cause
Effect
Message
Action
Beeps/Vib
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 169
Sensor/Transmitter and pump are not
communicating.
Insulin deliveries continue. CGM session stops.
Displayed each time pump is awakened until
confirmed.
Press
to confirm. Contact Customer Service for
assistance.
User selected (under CGM Setup menu options),
every 3 minutes until confirmed. No progression.
Transmitter battery power is low.
Insulin deliveries continue. CGM continues but with
possible data gaps.
Displayed each time the pump is awakened until
confirmed.
Press
to confirm. You must replace the entire
Transmitter. See Chapter 9 in Section II for more
information.
User selected (under CGM Setup menu options),
every 3 minutes until confirmed. No progression.
169
Dec 2014
CGM Warning
CGM Failure,
Insulin
Delivery
Continues,
Call Service
XXX-XXXXXXXX
XXXXXXXXXX
Confirm
CGM Warning
Replace CGM
Transmitter
Low Battery
Confirm
A279 of A312
11/14/14 2:53 PM
Need help?
Do you have a question about the g4 platinum and is the answer not in the manual?