Dexcom g4 platinum Quick Start Manual page 177

Sensor and transmitter
Hide thumbs Also See for g4 platinum:
Table of Contents

Advertisement

ANIMAS® VIBE™ SYSTEM
C H AP TE R 6 - V I EW IN G CGM IN F OR M A T ION O N Y O U R PU M P
Missing CGM Readings –
12:34 PM
BG
IOB 5.54U
CGM Warning
Enter BG
Confirm
Follow the steps for entering a fingerstick BG value (see Chapter 5 in Section II). After entering a
fingerstick BG value, a CGM reading should immediately appear on the display, and CGM readings
should resume being updated every 5 minutes. If a CGM reading does not immediately appear, follow
any display prompts and refer to Chapter 10 in Section II to troubleshoot any warnings, alarms, or
alerts.
The
symbol will remain on CGM Data and CGM Trend screens until the calibration
update/recalibration was successful. Your pump will then begin/resume displaying CGM readings.
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 149
appears in place of current CGM reading
Any time you see the
symbol on either the CGM Data or CGM Trend
screens, your pump requires that you enter a fingerstick BG value for a
calibration update or recalibration. This is because the Sensor needs to
recalibrate based on your current BG level.
A CGM Warning screen will also appear to remind you to enter a fingerstick
BG value.
149
Dec 2014
A259 of A312
11/14/14 2:53 PM