ANIMAS® VIBE™ SYSTEM
C H AP TE R 11 - P U M P S AF ET Y S Y ST EM A N D A L A RM S
Alerts, Warnings and Alarms
Warning: Exceeds Max Basal
Cause
Effect
Message
Action
Beeps/Vib
Warning: Delivery Canceled due to Low Cartridge
Cause
Effect
Message
Action
Beeps/Vib
Warning: No Prime, No Delivery
Cause
Effect
Message
Action
Beeps/Vib
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 88
Basal delivery rate (or Temp Basal delivery) exceeds
user-set maximum.
Basal delivery stops.
Displayed when awakened (by basal delivery attempt
every 3 minutes or manually) until confirmed.
Press
to confirm. Reprogram Max Basal amount in
the Setup Advanced menu (or reprogram Temp Basal).
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vib within one hour.
Basal or Bolus delivery exceeds insulin remaining in
cartridge.
Basal or Bolus delivery stopped.
Once per occurrence and each time awakened until
confirmed.
Press
to confirm. Replace with full cartridge.
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vib within one hour.
Pump is not primed.
All insulin deliveries stop.
Every 3 minutes or when awakened manually.
Press
to confirm. Disconnect, reprime.
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vib within one hour.
88
Dec 2014
Warning
Exceeds
max basal
XX.XX U/Hr.
No basal
delivery.
Confirm
Warning
Delivery
canceled
due to
low
cartridge.
Confirm
Warning
Pump is
not primed.
No delivery.
Confirm
A196 of A312
11/14/14 2:53 PM