ANIMAS® VIBE™ SYSTEM
C H AP TE R 5 - US I N G B AS AL PR O GR AM F EA TU R ES
Canceling a Temporary Basal Program
1.
From the BASAL MENU, select "TEMP BASAL". Press
2.
Details of the current active Temp Basal program will be displayed. Scroll
up to "CANCEL" and press
Your previously active basal program will be activated and the Home screen
will be displayed to show the current rate per hour of the active basal
program.
NOTE: If you Suspend your pump while a Temp Basal program is active, the
Temp Basal will be canceled and an Alert screen will notify you that the Temp
Basal program has been canceled. Your active basal program/rate will resume
once insulin delivery resumes. This Alert is displayed once and gives an
audible tone once (if you turned on Alert sounds in Setup). Temp Basal is also
canceled when you change the battery and/or prime.
As an example, let's say you set a Temp Basal that would reduce your basal rate by 40% for the next
2 hours.
If your current active basal rate is 1.000 U/Hr, your Temp Basal rate setting will adjust the rate to
0.600 U/Hr (1.000 U/Hr - 0.400 U/Hr).
During the 2 hour period you decide to suspend your pump at which time all insulin delivery will stop,
and the Temp Basal in effect will be canceled. Upon resuming insulin delivery, your basal rate will
resume to your active basal program/rate of 1.000 U/hr.
m
WARNING:
• ALWAYS review changes in your pump settings with your HCP to make sure they are correct.
Incorrect settings can result in under delivery or over delivery of insulin.
• ALWAYS use a fingerstick BG for treatment decisions, such as how much insulin you should
take. The Sensor and Transmitter do not replace a BG meter and BG values may differ from
Sensor glucose readings. Using glucose readings from the Sensor and Transmitter to make
treatment decisions can result in serious injury or death.
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 33
.
33
Dec 2014
.
A1-WEEKDAY
TEMP ACTIVE
= 0.405 U/Hr
CANCEL
Main Menu
Pump Suspended
Temp Basal &
Combo Bolus
0.675 U/Hr
-40%
Time left
1.2 Hr
Alert
If active,
have been
canceled.
A141 of A312
11/14/14 2:52 PM