C H AP TE R 11 - P U M P S AF ET Y S Y ST EM A N D A L A RM S
Alerts, Warnings and Alarms
Warning: Suspend
Cause
Effect
Message
Action
Beeps/Vib
Warning: No Cartridge Detected, Deliveries Disabled
Cause
Effect
Message
Action
Beeps/Vib
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 85
Pump suspended manually.
All insulin deliveries stop.
Displayed each time pump is awakened. Then displayed
once every 3 minutes upon confirmation, and once
every 15 minutes if not confirmed, until action is taken.
Press
to confirm. Resume delivery.
User selected, once every 3 minutes if confirmed, once
every 15 minutes if not. No progression. Sweep/vib
within one hour only if bolus or basal delivery was in
progress when pump was suspended.
No cartridge detected after "Load cartridge" step during
Prime/Rewind.
No insulin deliveries.
Displayed when manually awakened until confirmed.
Press
to confirm. Be sure Prime/Rewind sequence is
completed with cartridge properly in place.
User selected, once every 3 minutes. No progression if
confirmed each time displayed. Sweep/vib within one
hour if not confirmed.
85
Dec 2014
Warning
No delivery.
Pump is
suspended.
Confirm
Warning
No cartridge
detected.
Delivery
disabled.
Confirm
A193 of A312