ANIMAS® VIBE™ SYSTEM
C H AP TE R 9 - AD V AN C ED F EA TU R E S / SE TU P A N D A CT IV AT I ON
Setup Advanced Screen 4 – Language Setup, Display Timeout, Contrast and
Battery Type
You can customize how information will be displayed on your pump, and what type of battery you will use.
This screen allows you to:
• Select a different language
• Set the length of time your display stays on before timing out to save
battery life
˚
15, 30, 45 or 60 seconds
• Select a contrast setting
• Select Lithium (recommended) or Alkaline battery type. If you select the wrong battery type, the alarm
screens may not function properly. You can also change the battery type on the VERIFY screen when
you insert a new battery.
1.
From the SETUP ADV 4 screen, scroll to the desired field.
2.
Press
to change to flashing cursor for Edit mode.
3.
Use the
/
4.
Press
when setting is made.
5.
Scroll to "
Advanced Setup screens, or "Home" to go to the Home screen.
6.
Press
.
Contrast Button/CGM Shortcut
Pressing the
contrast levels: Dim, Default and Bright. To preserve battery life, your pump display will Auto-dim
when no pump function button (any button other than the Contrast or Audio Bolus buttons) has been
pressed for half the period you set for the display to time out under Advanced Setup screen 4. For
example, if you set the display time out for 60 seconds, the pump will Auto-dim with no button pushes
in a 30 second period. While in Auto-dim mode, you can restore the default contrast level you have set
by pressing the
restore the default contrast level as well as perform the function of the button. If in Call Service Alarm
mode, you must use the
To return contrast setting to original factory default, press the
time. When the word "Contrast" is displayed on the screen, press any button to return to the default
contrast setting.
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 54
buttons to change to desired setting.
", "
", or "Home" at the bottom of the screen. Select "
button on the top of your pump adjusts the contrast of your display. There are three
button on top of your pump. Pressing a function button while in Auto-dim mode will
button to restore the default contrast level.
" or "
button and
54
Dec 2014
SETUP ADV 4
Language
ENGLISH
Display Timeout
in 60 secs
Contrast
8
Battery
Lith
Home
" to go to other
button at the same
A162 of A312
11/14/14 2:52 PM
Need help?
Do you have a question about the g4 platinum and is the answer not in the manual?