C H AP TE R 3 - GE T TI N G Y O U R PU M P R EA D Y
13.
Press the
screen.
NOTE: This step is not necessary for needle sets.
14.
Use the
Instructions for Use included with your infusion set for details on how much insulin is required to fill
the cannula. Press
15.
Press
to fill the cannula.
NOTE: The maximum Fill Cannula amount is 1U at a time.
Be sure the infusion set is not connected to your body until the prime is
complete.
If your pump is suspended, the screen will alert you with the ezPrime "Pump
suspended" screen. You must resume delivery of your pump in order to
complete the Priming function. Scroll to "Home" on the ezPrime screen and
press
to display the MAIN MENU. Select "Suspnd/Resum" and follow the
steps for resuming insulin delivery. See Chapter 6 in Section I.
NOTE: The Fill Cannula step is not required for your pump to operate. For example, when you prime
your pump after a battery change and you are not inserting a new infusion set, this step is not
necessary. Filling the cannula when not necessary can result in unwanted delivery of insulin.
Selecting the Infusion Site and Inserting the Infusion Set
Your HCP will review appropriate site selections and techniques for insertion based on your body type.
Refer to the Instructions for Use included with your infusion set for proper insertion guidelines.
m
WARNING: AVOID infusion sites on skin areas with tattoos, or areas with rough patches or
scarring from your pump or insulin injections. These skin areas can cause redness, irritation,
swelling, infection, and not allow for the intended amount of insulin delivery.
m
CAUTION: If you are using the Dexcom G4 PLATINUM Sensor and Transmitter with your
pump, AVOID injecting insulin or placing an insulin pump infusion set within 3 inches
(7.62 centimeters) of the Sensor. Insulin delivery within 3 inches (7.62 centimeters) of the
Sensor can cause inaccurate Sensor glucose readings.
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 23
button to display the FILL CANNULA
/
buttons to enter the amount of insulin needed to fill the cannula. Refer to the
so that "Go" is highlighted.
ezPrime
Rewind
Load Cart
Prime
Fill Cannula
Main Menu
23
Dec 2014
FILL CANNULA
Amount
U
0.00
Go
Main Menu
ezPrime
Pump
suspended.
Priming
is
disabled.
Rewind
Home
A131 of A312
11/14/14 2:52 PM