ANIMAS® VIBE™ SYSTEM
C H AP TE R 11 - P U M P S AF ET Y S Y ST EM A N D A L A RM S
Alerts, Warnings and Alarms
Alarm: Call Service
Cause
Effect
Message
Action
Beeps/Vib
NOTE: Some Call Service Alarms have a unique sound/vibration sequence
and cannot be silenced by pressing
For these Alarms the usual progression is replaced by 3 chirps/vib repeated
every 9 minutes for the first half hour. This is followed by 4 long tones/vib
after that.
Alarm: Auto-Off
Cause
Effect
Message
Action
Beeps/Vib
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 91
Hardware or software problem detected.
All insulin deliveries suspended.
Continuous until battery is removed.
Remove pump battery to silence the alarm, or
press
to silence alarm for 30 minutes (alarm can
only be silenced once). Contact Customer Service.
Fixed beep/vib tone. If not confirmed, progresses to
sweep/vib within one hour.
.
Pump suspended because there were no button
presses on the pump in the time period set by the
user for the AUTO-OFF Alarm feature.
All insulin deliveries stop.
Continuous until confirmed.
Press
to confirm. Adjust time period in the Setup
Advanced menu.
User-selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vib within one hour.
(Once confirmed, No Prime warning triggered, see
No Prime Warning screen in this chapter.)
91
Dec 2014
ALARM
CALL SERVICE
No delivery.
XX-XXXXX
Remove battery
to silence
the alarm.
ALARM
AUTO-OFF
No delivery.
No button
presses in last
XX hours.
Confirm
A199 of A312
11/14/14 2:53 PM