Dexcom G4 Platinum User Manual
Hide thumbs Also See for G4 Platinum:
Table of Contents

Advertisement

Quick Links

CONTINUOUS GLUCOSE
MONITORING SYSTEM
USER'S GUIDE

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the G4 Platinum and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Dexcom G4 Platinum

  • Page 1 CONTINUOUS GLUCOSE MONITORING SYSTEM USER’S GUIDE...
  • Page 3 CONTINUOUS GLUCOSE MONITORING SYSTEM...
  • Page 4 Other Notes: © 2014 Dexcom, Inc. All rights reserved. Dexcom, Dexcom G4, Dexcom G4 PLATINUM, Dexcom Studio and SEVEN are either registered trademarks or trademarks of Dexcom, Inc. in the United States and/or other countries. All other product or company names that may be mentioned in this publication are tradenames, trademarks or registered trademarks of their respective owners.
  • Page 5 DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING (CGM) SYSTEM Dexcom G4 PLATINUM Receiver Dexcom G4 PLATINUM Sensor Dexcom G4 PLATINUM Transmitter SYSTEM CONTENTS: Reorder only the Dexcom G4 PLATINUM System components listed below: • sensor • receiver case • transmitter • user’s guide • receiver •...
  • Page 6: Dexcom G4 Platinum Continuous Glucose Monitoring System

    Some people have found this to be an effective method of product training. Please review the tutorial on the disc and discuss with your healthcare professional if the Dexcom G4 PLATINUM training tutorial is an appropriate training option for you.
  • Page 7: Indications For Use

    (age 2 and older) with diabetes. The system is intended for use by patients at home and in healthcare facilities. The Dexcom G4 PLATINUM System is indicated for use as an adjunctive device to complement, not replace, information obtained from standard home glucose monitoring devices.
  • Page 8 WARNINGS • Do not use the Dexcom G4 PLATINUM CGM System until after you have been trained or viewed the training materials included with your CGM system. • The blood glucose value from your blood glucose meter should be used for treatment decisions, such as how much insulin you should take.
  • Page 9 Sensors should not be stored in a freezer. • Keep the USB port cover on the receiver closed whenever the USB cable is not attached. The receiver may malfunction if water gets into the USB port. Dexcom G4 PLATINUM User’s Guide...
  • Page 10: Table Of Contents

    2.2.2 Setting Your Receiver Time and Date 2.2.3 Entering Your Transmitter ID 2.2.4 Setting Your Receiver Time Format 2.2.5 Selecting Your Receiver Language 2.3 Checking Information About Your Dexcom G4 PLATINUM System 2.4 Transmitter and Receiver Communication CHAPTER 3: INSERTING A SENSOR AND STARTING A SENSOR SESSION 3.1 Before You Start...
  • Page 11 CHAPTER 4: CALIBRATING YOUR DEXCOM G4 PLATINUM SYSTEM 4.1 Calibration Overview 4.2 How to Calibrate 4.3 Startup Calibration 4.4 12 Hour Calibration Update 4.5 Other Reasons You May Need to Calibrate CHAPTER 5: SENSOR GLUCOSE READINGS AND TRENDS 5.1 Sensor Glucose Readings 5.2 Rate of Change Arrows...
  • Page 12 9.8.2 Receiver Error Code 9.8.3 Low Transmitter Battery 9.8.4 Transmitter Failed Error Code 9.9 Out of Range/No Antenna 9.10 Alerts Not Working CHAPTER 10: TAKING CARE OF YOUR DEXCOM G4 PLATINUM SYSTEM 10.1 Maintenance 10.2 Storage 10.3 Cleaning and Disinfection 10.4 Product Disposal...
  • Page 13: Chapter 1: The Dexcom G4 Platinum System Basics

    THE DEXCOM G4 PLATINUM SYSTEM BASICS Dexcom G4 PLATINUM User’s Guide...
  • Page 14 Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 15: Sensor Overview

    The transmitter battery will last at least 6 months. Once you see the transmitter low battery screen shown on the right, replace the transmitter as soon as possible. Your transmitter battery may drain as quickly as one week after this alert appears. Dexcom G4 PLATINUM User’s Guide...
  • Page 16: Receiver Overview

    Antenna Trend Arrow High Glucose Alert Setting Glucose Values Glucose Plot of most Target Range recent sensor glucose reading Low Glucose Current Time Alert Setting Trend Graph Hours EXAMPLE: 3-Hour Trend Graph Screen Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 17 Your receiver and transmitter are wirelessly paired together to communicate securely and only with each other. You will need a commercially available blood glucose meter to use with your system. UP button RIGHT button SELECT button LEFT button DOWN button Dexcom G4 PLATINUM User’s Guide...
  • Page 18: Charging Your Receiver Battery

    • an AC power outlet • a personal computer with Windows operating system (to charge your receiver from your PC, Dexcom Studio must be installed. For system requirements and more information, see the Dexcom website (www.dexcom.com) or the Dexcom Studio Software User’s Guide.) It takes about 3 hours to fully charge an empty battery with the wall charger, and about 5 hours when the receiver is connected to a computer.
  • Page 19 • Press down firmly with your thumb when sliding open the USB port cover. 5. The battery charging screen will appear on the receiver. 6. After a few seconds the trend graph will reappear with the battery charging symbol ( ) shown in the upper left corner. Dexcom G4 PLATINUM User’s Guide...
  • Page 20: Charging Your Receiver Battery From Awindows Compatible Computer

    1.4.2 CHARGING YOUR RECEIVER BATTERY FROM A WINDOWS COMPATIBLE COMPUTER The Dexcom Studio software must be installed to charge your receiver from a computer. 1. Plug the included USB cable into your computer. 2. Plug other end of the USB cable into the receiver.
  • Page 21: Knowing Your Receiver Is Charged

    After 3 days you will be prompted to reset your receiver time and date (see Chapter 2, Section 2.2, The Settings Menu). • Only use the Dexcom battery charger provided in the receiver kit. Do not use any other battery charger.
  • Page 22 Settings To change the time, date and transmitter ID, to look up Chapter 2 your Dexcom G4 PLATINUM System hardware and software version numbers, to view transmitter battery status, last calibration value, sensor insertion time. Shutdown...
  • Page 23: Chapter 2: Dexcom G4 Platinum System Setup

    DEXCOM G4 PLATINUM SYSTEM SETUP Dexcom G4 PLATINUM User’s Guide...
  • Page 24 Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 25: The Setup Wizard

    This chapter will help you when you first set up your Dexcom G4 PLATINUM Continuous Glucose Monitoring System. Read this chapter before you start. THE SETUP WIZARD 1. Before setting up the receiver remove the transmitter from its packaging. It takes 10 minutes for the transmitter to turn on after it is removed from its packaging.
  • Page 26 Numbers list on page 4. • If you need to change the language, time format, time, date or transmitter ID after you have completed the Setup Wizard see Section 2.2, The Settings Menu. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 27 SELECT button to confirm your preferred alert value. The Setup Wizard is now complete. To begin using your Dexcom G4 PLATINUM CGM System you will need to insert a sensor (see Chapter 3, Inserting a Sensor and Starting a Sensor Session).
  • Page 28: The Settings Menu

    RIGHT button to move to the next value. The date format is YYYY/MM/DD. After choosing the time, press the SELECT button. You will return to the settings menu. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 29: Entering Your Transmitter Id

    Continue repeating steps a and b until you have the transmitter ID displayed. c. After entering the fifth number or letter, press the SELECT button. You will return to the settings menu. Dexcom G4 PLATINUM User’s Guide...
  • Page 30: Setting Your Receiver Time Format

    “Language” and press the SELECT button. 2. Press the DOWN button to scroll and highlight your desired language. 3. Press the SELECT button to set the highlighted language. You will return to the settings menu. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 31: Checking Information About Your Dexcom G4 Platinum System

    CHECKING INFORMATION ABOUT YOUR DEXCOM G4 PLATINUM SYSTEM You can check your receiver for information about your CGM at any time. 1. From the Settings menu, press the DOWN button to scroll to “Device Info” and press the SELECT button.
  • Page 32: Transmitter And Receiver Communication

    The Dexcom G4 PLATINUM Sensor, Transmitter, and Receiver must be removed prior to Magnetic Resonance Imaging (MRI), CT scan, or diathermy treatment. The Dexcom G4 PLATINUM System has not been tested during MRI, CT scans or with diathermy treatment, and it is unknown if there are safety or performance issues.
  • Page 33: Chapter 3: Inserting A Sensor And Starting A Sensor Session

    INSERTING A SENSOR AND STARTING A SENSOR SESSION Dexcom G4 PLATINUM User’s Guide...
  • Page 34 Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 35: Before You Start

    If you experience a broken sensor, please report this to your local distributor. For patients undergoing an MRI with a retained wire broken off from a Dexcom G4 PLATINUM Sensor, in-vitro MRI testing did not detect any safety hazards.
  • Page 36: Removing The Sensor From Its Packaging

    Sensor pod Release tab REMOVING THE SENSOR FROM ITS PACKAGING The sensor is sterile in its unopened, undamaged package. Do not use any sensor if its sterile package has been previously damaged or opened. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 37: Choosing An Insertion Site

    (front of body, option A) or the upper buttocks (back of body, option B). No other sensor insertion sites have been tested. Front of body, sensor site option A Back of body, (ages 2 and older) sensor site option B (ages 2-17 only) Dexcom G4 PLATINUM User’s Guide...
  • Page 38: Placing The Sensor

    3. Place the sensor flat on your selected area. Make sure the sensor is placed in the same direction shown in the picture below. You should not place the sensor pointing in the up or down direction. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 39: Sensor Insertion

    Dexcom CGM. SENSOR INSERTION Once you have placed the applicator on your skin and removed the safety lock, you are ready to insert the sensor. To insert your sensor follow these steps: Dexcom G4 PLATINUM User’s Guide...
  • Page 40 After this step, only the sensor pod will be left on your body. • Make sure the transmitter latch is down Release the applicator barrel (against your body) to remove the applicator barrel. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 41: Transmitter Attachment

    You can release your pinch on the adhesive edge at this time. c. Make sure the transmitter is secure by sliding your first and second fingers under the sensor pod wings and press down on the transmitter with your thumb. Dexcom G4 PLATINUM User’s Guide...
  • Page 42: Starting A Sensor Session

    The start sensor processing screen will appear to let you know your sensor session has been activated and your 2-hour startup has begun. 5. Your receiver will then return to the 3-hour trend graph. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 43: Sensor Startup Period

    See the pictures below for an example of what this looks like. During the countdown period, you will not get sensor glucose readings, alerts and alarms. Start Ready for [0-24] [24-48] [48-72] [72-96] minutes minutes minutes minutes calibration Dexcom G4 PLATINUM User’s Guide...
  • Page 44: Taping The Sensor Pod

    At the end of the 2-hour startup period the receiver will let you know that it is time to calibrate your sensor. Chapter 4, Calibrating Your Dexcom G4 PLATINUM System, tells you how to calibrate your sensor.
  • Page 45: The Dexcom G4 Platinum System And Water

    Your sensor session will stop 7 days after you started the sensor session. 3.11 THE DEXCOM G4 PLATINUM SYSTEM AND WATER Your sensor is water resistant when showering, bathing, or swimming if the transmitter is fully snapped in. The sensor has been tested to be water resistant when submerged for up to 2.44 meters for a maximum of 24 hours.
  • Page 46 THIS PAGE LEFT INTENTIONALLY BLANK Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 47: Chapter 4: Calibrating Your Dexcom G4 Platinum System

    CALIBRATING YOUR DEXCOM G4 PLATINUM SYSTEM Dexcom G4 PLATINUM User’s Guide...
  • Page 48 Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 49: Calibration Overview

    The Dexcom G4 PLATINUM System requires you to calibrate the sensor glucose readings to your blood glucose meter. Taking acetaminophen (paracetamol) containing products (such as Tylenol) while wearing the sensor may falsely raise your sensor glucose readings. The level of inaccuracy depends on the amount of acetaminophen (paracetamol) active in your body.
  • Page 50: How To Calibrate

    DOWN button to scroll until you highlight “Enter BG.” 4. Press the SELECT button to choose this option. You will see a screen with a blood drop and a number in mmol/L units. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 51 “time out” and no blood glucose value will be recorded for calibration. 7. The “Enter BG” processing screen will appear to let you know the blood glucose value is being processed for calibration. Dexcom G4 PLATINUM User’s Guide...
  • Page 52 • The accuracy of the blood glucose meter measurement used for calibration may affect the accuracy of sensor glucose readings. • Do not calibrate your Dexcom G4 PLATINUM System when you have acetaminophen (paracetamol) active in your body. Taking acetaminophen (paracetamol) containing products (such as Tylenol) while wearing the sensor may falsely raise your sensor glucose readings.
  • Page 53: Startup Calibration

    12 hours, the receiver will ask you to enter a blood glucose value to update its calibration. The following steps show you how to enter this calibration. Your sensor glucose readings may be inaccurate if you calibrate less than every 12 hours. Dexcom G4 PLATINUM User’s Guide...
  • Page 54: Other Reasons You May Need To Calibrate

    9.2.1, Types of Calibration Prompts. Calibration prompt HELPFUL HINT: If you see these screens they are indicating calibration errors (see Chapter 9, Section 9.3, Calibration Error Troubleshooting). 15 minute calibration 1 hour calibration error screen error screen Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 55: Chapter 5: Sensor Glucose Readings And Trends

    SENSOR GLUCOSE READINGS AND TRENDS Dexcom G4 PLATINUM User’s Guide...
  • Page 56 Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 57 HELPFUL HINTS: • The greatest benefit you receive from using your Dexcom G4 PLATINUM System will come from the trending information. It is important that you focus on the trends and rate of change on your receiver, rather than the exact glucose reading.
  • Page 58: Sensor Glucose Readings

    • If your low glucose alert is not set and your glucose is 3.1 mmol/L or lower, your glucose value will be red. • If your sensor glucose readings are in between your high and low glucose alert settings, the glucose value will be white. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 59 6 hours of sensor glucose readings. (Scroll down from the 6-hour graph to reach the 12-hour graph) 12-Hour Trend Graph: The 12-hour trend graph shows your current glucose reading and the last 12 hours of sensor glucose readings. Dexcom G4 PLATINUM User’s Guide...
  • Page 60: Rate Of Change Arrows

    HELPFUL HINT: • Do not overreact to the rate of change arrows. Consider recent insulin dosing, activity, food intake, your overall trend graph and your blood glucose value before taking action. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 61 If it continued falling at this rate, your glucose could decrease more than 2.5 mmol/L in 15 minutes. No rate of change information: The receiver cannot calculate how fast your glucose is rising or falling at this time. arrow Dexcom G4 PLATINUM User’s Guide...
  • Page 62: Glucose Status Area Symbols

    (see Chapter 4, Section 4.2, How to Calibration Calibrate). Needed This icon means you need to enter one more blood glucose value in order to calibrate the system and start receiving sensor glucose readings. Additional Calibration Needed Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 63 This icon means the receiver has detected a potential problem with the sensor signal. You should wait about 30 minutes for more prompts. Do not enter any blood glucose values when you see this symbol. Wait Dexcom G4 PLATINUM User’s Guide...
  • Page 64 THIS PAGE LEFT INTENTIONALLY BLANK Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 65: Chapter 6: Alerts, Alarms And Alert Profiles

    ALERTS, ALARMS AND ALERT PROFILES Dexcom G4 PLATINUM User’s Guide...
  • Page 66 Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 67: Setting Your Alerts

    CGM System’s many alerts and alarms and how to customize them. SETTING YOUR ALERTS 6.1.1 DEFAULT ALERT/ALARM SETTINGS When you receive your Dexcom G4 PLATINUM CGM System the following alerts and alarms are preset on your receiver (see table below). What will I see What does this What...
  • Page 68 • Your receiver may alert you at other times you need to take action such as low battery, failed sensor, etc. See Chapter 12, Appendix, for a detailed list of these other alerts. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 69: Glucose Alerts

    6.1.2 GLUCOSE ALERTS The Dexcom G4 PLATINUM System lets you create your own personal settings for how you want the receiver to tell you what is going on. The low and high glucose alerts will tell you when your sensor glucose readings are outside your target glucose range.
  • Page 70: Low Glucose Alert

    • When you have both your high and low alerts turned on there will be a gray zone on your trend graphs indicating your target range. • If your high or low alert is turned off this gray zone will not appear. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 71: Low Glucose Alarm

    6.1.5 LOW GLUCOSE ALARM The Dexcom G4 PLATINUM System also has a fixed low alarm set at 3.1 mmol/L. This alarm is a feature in addition to your personal low and high glucose alerts. You cannot change or turn off this alarm or its re-alarm settings.
  • Page 72: High And Low Glucose Alerts

    • Your high glucose alert value can be set between 6.7 and 22.2 mmol/L in 0.1 mmol/L increments. • Your low glucose alert value can be set between 3.3 and 5.5 mmol/L in 0.1 mmol/L increments. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 73: Advanced Alerts

    3. Press the DOWN button to highlight “Alerts” and press the SELECT button. 4. From the alerts menu, press the DOWN button to select “Advanced” and press the SELECT button. Dexcom G4 PLATINUM User’s Guide...
  • Page 74: Setting A Snooze Time For Your High And Low Glucose Alerts

    Press the SELECT button. If you set the amount of time to zero there will be no re-alerts. 3. When finished, press the LEFT button to return to the alerts menu. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 75: Rise And Fall Glucose Rate Alerts

    If you set your rise rate to 0.17 mmol/L per minute and your sensor glucose readings rise at this rate or faster, the “RISING double arrow” screen will appear and the receiver will vibrate or beep according to your profile settings. Rapid rise alert Dexcom G4 PLATINUM User’s Guide...
  • Page 76 (0.11 mmol/L or more per minute) or “0.17 mmol/L/min” (0.17 mmol/L or more per minute). Press the SELECT button. 5. When finished, press the LEFT button to return to the alerts menu. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 77: Setting The Out Of Range Alert

    “On.” If you do not want to receive out of range alerts press the SELECT button to choose “Off.” 3. Next, on the out of range menu, press the DOWN button to choose “Time” and press the SELECT button. Dexcom G4 PLATINUM User’s Guide...
  • Page 78: Alert Profiles

    ALERT PROFILES The Dexcom G4 PLATINUM System allows you to set profiles to choose the way you want your alerts to behave. This customizable feature is found under the Profiles option on the Main Menu shown to the right.
  • Page 79 • If the alert is confirmed and your sensor glucose readings continue to be at or below 3.1 mmol/L your system will repeat the alert sequence above in 30 minutes. To confirm any alert press the SELECT button. Dexcom G4 PLATINUM User’s Guide...
  • Page 80: Alert Profile Options

    The “Try It” feature is available under the profiles menu and allows you to hear an example of the alert sequence and sound for each individual alert and alarm. Once you hear the sounds, it is easier to understand the alert profile options. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 81: Alert Profile Details

    All Other 1 long vibrate 1 long 1 long 1 long vibrate 1 long vibrate + 1 vibrate + 1 vibrate + + 1 short medium beep Alerts low beep 1 medium melody beep Dexcom G4 PLATINUM User’s Guide...
  • Page 82 The profiles menu will appear. 4. Press the DOWN button to scroll to the profile you want to set and press the SELECT button. 5. When finished, press the LEFT button to return to the main menu. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 83: Chapter 7: Events

    EVENTS Dexcom G4 PLATINUM User’s Guide...
  • Page 84 Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 85: Events

    Dexcom Studio software. EVENTS HELPFUL HINT: • Event markers can be downloaded and viewed in the Dexcom Studio software but cannot be viewed on your receiver. 7.1.1 SELECTING AN EVENT 1. Press the SELECT button to turn on the receiver.
  • Page 86: Setting The Date And Time For An Event

    RIGHT button to move to the next value. After adjusting the time, press the SELECT button. 2. You will advance to the event confirmation screen. Press the SELECT button to confirm the entry. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 87: Carbohydrates

    4. You will advance to the event confirmation screen. Press the LEFT RIGHT button to choose either “OK” to confirm or “Cancel” to discard this entry, and then press the SELECT button. You will return to the Events menu. Dexcom G4 PLATINUM User’s Guide...
  • Page 88: Insulin

    4. You will advance to the event confirmation screen. Press the LEFT RIGHT button to choose either “OK” to confirm this entry or “Cancel” to discard this entry, and then press the SELECT button. You will return to the Events menu. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 89: Exercise

    5. You will advance to the event confirmation screen. Press the LEFT RIGHT button to choose either “OK” to confirm this entry or “Cancel” to discard this entry, and then press the SELECT button. You will return to the events menu. Dexcom G4 PLATINUM User’s Guide...
  • Page 90: Health

    You will return to the events menu. DEXCOM STUDIO SOFTWARE The Dexcom Studio software is an optional component of your system. This software allows you to view trends, track patterns and create customizable charts to display your glucose trends.
  • Page 91: Chapter 8: Completing A Sensor Session

    chapter eight COMPLETING A SENSOR SESSION...
  • Page 93 (sensor and applicator). • In some cases your sensor session may end before you have completed a full 7-day period. If this happens, see Chapter 9, Section 9.6, Sensor Shut-Off Troubleshooting. Dexcom G4 PLATINUM User’s Guide...
  • Page 94: Automatic Sensor Shut-Off

    You will need to remove your sensor and insert a new sensor. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 95: Removing A Sensor

    1. Place the sensor pod on a table. 2. Hold the rounded edge of the safety lock. 3. Make sure the jagged edge of the safety lock is facing down (the direction away from the removal arrow) as shown below: Dexcom G4 PLATINUM User’s Guide...
  • Page 96 If you did not save the safety lock, you can simply use your fingers to spread out the tabs at the back of the sensor pod (end closest to the sensor pod tab wings). The transmitter will “pop” out of the sensor pod. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 97: Chapter 9: Troubleshooting

    TROUBLESHOOTING Dexcom G4 PLATINUM User’s Guide...
  • Page 98 Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 99: Sensor Insertion Troubleshooting

    This chapter provides helpful tips and instructions to resolve issues you may experience while using your Dexcom G4 PLATINUM CGM System. HELPFUL HINT: • If the troubleshooting steps in this chapter do not fix your issue, contact your local distributor.
  • Page 100: Calibration Troubleshooting

    • Make sure you have not taken any medications containing acetaminophen (paracetamol) (such as Tylenol). • Make sure your transmitter is fully snapped in to your sensor pod. • See Chapter 4, Calibration, for further information. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 101: Types Of Calibration Prompts

    The receiver will continue to display this prompt screen every 15 minutes until the receiver accepts the blood glucose values. Calibration prompt To clear this prompt, press the SELECT button. Dexcom G4 PLATINUM User’s Guide...
  • Page 102: Status Area Symbols During Calibration Prompts

    If you clear this screen by pressing the SELECT button, this symbol will be displayed in the status area. Wait 1 hour calibration error screen Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 103: System Glucose Error

    Place the transmitter on a clean, dry cloth and air dry for 2-3 minutes. Often the system can correct this problem and continue providing sensor glucose readings. However, if it has been at least 3 hours since your last glucose reading appeared, contact your local distributor. Dexcom G4 PLATINUM User’s Guide...
  • Page 104: Sensor Inaccuracies

    If one or more of your readings entered was outside of this range, the receiver will not calibrate. You will have to wait until your blood glucose is in this range to calibrate. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 105: Sensor Shut-Off Troubleshooting

    • Make sure nothing is rubbing the sensor pod (i.e. clothing, seat belts, etc.). • Make sure you have selected a good insertion site (see Chapter 3, Section 3.3, Choosing an Insertion Site). • Make sure your insertion site is clean and dry prior to sensor insertion. Dexcom G4 PLATINUM User’s Guide...
  • Page 106: Early Sensor Shut-Off - Sensor Failure

    - When you are not in an active sensor session, you will not see the “Stop Sensor” option on the main menu. • Remember to stop your sensor session if you remove your sensor before the end of the full 7 day period. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 107 3. The stop sensor processing screen will appear to let you know the sensor session is stopping. 4. Once the session has stopped, a red stoplight symbol ( ) will appear in the upper right corner of the trend graph. Dexcom G4 PLATINUM User’s Guide...
  • Page 108: Battery And Charger Troubleshooting

    BATTERY AND CHARGER TROUBLESHOOTING Only use the Dexcom cable and battery charger to charge your receiver. 1. Remember to charge your receiver battery as needed. The charge will last approximately 3 days, depending on how often you turn on your receiver, use the alerts, and enter events.
  • Page 109: Low Transmitter Battery

    • If you see the out of range symbol in the status area for more than 10 minutes, move your receiver and transmitter within 6 meters of each other. Wait 10 minutes and communication should be restored. Dexcom G4 PLATINUM User’s Guide...
  • Page 110: Alerts Are Not Working

    • If your receiver gets wet or is dropped you need to make sure the speakers and vibrate mode are still working. You can do this by using the Try It option in the profiles menu (see Chapter 6, Section 6.3.1, Alert Profile Options). Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 111: Chapter 10: Taking Care Of Your Dexcom G4 Platinum System

    TAKING CARE OF YOUR DEXCOM G4 PLATINUM SYSTEM Dexcom G4 PLATINUM User’s Guide...
  • Page 112 Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 113: Maintenance

    • Only use Dexcom-supplied parts (including cables and chargers). Use of non- Dexcom supplied parts may affect safety and performance. There is no repair service available for your Dexcom G4 PLATINUM CGM System. If you experience problems with your system contact your local distributor.
  • Page 114: Storage

    4. The transmitter is water resistant when snapped into the sensor pod, but do not soak the transmitter by itself in liquid. 5. Do not use soap, nail polish remover, or paint thinner. Only use isopropyl alcohol and water. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 115 5. Dry the unit with a clean, dry absorbent wipe. 6. Wipe the outside of the device with a 70% isopropyl alcohol wipe to remove any disinfectant residue. 7. Dry the unit with a clean, dry absorbent wipe. Dexcom G4 PLATINUM User’s Guide...
  • Page 116: Product Disposal

    11. Wash hands thoroughly. 10.4 PRODUCT DISPOSAL Consult your local waste management authorities for appropriate procedures for disposal of devices containing electronic waste (transmitter and receiver) or blood contacting components (sensor and applicator). Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 117: Chapter 11: Technical Information

    TECHNICAL INFORMATION Dexcom G4 PLATINUM User’s Guide...
  • Page 118 Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 119: Product Specifications

    Sterilization Sterile by radiation Transmitter The Dexcom G4 PLATINUM Transmitter (Model 9438-01 and Model 9438-05) has been certified by Nemko Canada to electrical safety requirements of CSA C22.2 No. IEC 60601-1(ed. 3) and IEC 60601-1-11(ed. 1), under Certificate of Compliance NO67180/M1.
  • Page 120 1.25 mW EIRP Modulation Minimum Shift Key Data Rate 49.987 Kbits/Sec Total Packet 224 bits 4.48 ms every 5 minutes at each of the Transmit Duty Cycle four TX frequencies. Data Detection Range Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 121 (50/60 Hz) of a typical location in Magnetic Field a typical commercial or IEC 61000-4-8 hospital environment Note: U is the a.c. mains voltage prior to application of the test level. Dexcom G4 PLATINUM User’s Guide...
  • Page 122 Temperature: 0° C - 45° C Conditions Humidity: 10% - 95% RH Operating Altitude -152 to 3657 meters Moisture Protection IP22: vertically falling drops Limited Warranty 1 year Control Classification Class II equipment Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 123 (50/60 Hz) of a typical location in a Magnetic Field typical commercial or hospital environment. IEC 61000-4-8 Note: U is the a.c. mains voltage prior to application of the test level. Dexcom G4 PLATINUM User’s Guide...
  • Page 124 Guidance and Manufacturer’s Declaration – Electromagnetic Immunity The Dexcom G4 PLATINUM System is intended for use in the electromagnetic environment specified below. The customer or the user of the Dexcom G4 PLATINUM System should ensure that it is used in such an environment.
  • Page 125 To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Dexcom G4 PLATINUM System is used exceeds the applicable RF compliance level above, the Dexcom G4 PLATINUM System should be observed to verify normal operation.
  • Page 126 Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The Dexcom G4 PLATINUM System is intended for use in the electromagnetic environment specified below. The customer or the user of the Dexcom G4 PLATINUM System should ensure that it is used in such an environment.
  • Page 127 Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Dexcom G4 PLATINUM User’s Guide...
  • Page 128 USB charging/download cable for charging using an AC power outlet. Power Supply/Charger Dexcom P/N MT21255 Class Input: AC Input 100-240 Vac, 50/60Hz, 0.2A, 0.2A rms at 100Vac DC Output: 5V DC, 1A (5.0 Watts) Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 129: Device Performance Characteristics

    Dexcom G4 PLATINUM System performs. The Dexcom G4 PLATINUM System (the System) uses a glucose sensor to continuously measure and monitor your glucose levels. The sensor is “calibrated” using a commercially available blood glucose meter; and once calibrated the System reports glucose readings up to every 5 minutes.
  • Page 130 Number Percent Percent Percent Range of paired within 20/20% within 30/30% within 40/40% mg/dL CGM-YSI (mmol/L) Overall 9152 40-80 1351 (2.2-4.4) NOTE: CGM readings are within 40 to 400 mg/dL (2.2-22.2 mmol/L), inclusive. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 131 System reads about 9.8% or less different than YSI blood glucose values. The mean percent difference overall is 2.9%, and the median percent difference overall is 1.7%. This shows that the system is biased slightly higher than compared to the YSI. Dexcom G4 PLATINUM User’s Guide...
  • Page 132 Hypoglycemic Alert Evaluation – 15 Minutes Before and After Hypoglycemic Hypoglycemia Hypoglycemia True Alert False Alert Alert Level Detection Missed Detection Rate Rate mg/dL (mmol/L) Rate Rate 55 (3.1) 60 (3.3) 70 (3.9) 80 (4.4) 90 (5.0) 100 (5.6) Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 133 The Hypoglycemia Missed Detection Rate is the % of time the blood glucose was at or below the alert setting, but the device did not alarm within 15 or 30 minutes before or after the blood glucose was at or below the alert setting. Dexcom G4 PLATINUM User’s Guide...
  • Page 134 If your alarm soulds, you can expect your blood sugar to be at or above 200 mg/dL (11.1 mmol/L) approximately 8% of the time within the 15 minute period before or after your alarm sounds. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 135 Absolute within within within of Paired Wear Percent Percent 20/20% 30/30% 40/40% CGM-YSI Differences (%) Differences (%) YSI* YSI* YSI* Day 1 3023 16.7% 13.2% Day 4 3108 11.4% 8.2% Day 7 3021 11.9% 8.9% Dexcom G4 PLATINUM User’s Guide...
  • Page 136 System Agreement to SMBG within CGM Glucose Ranges Percent Percent Percent CGM Glucose Number of within within within Ranges mg/dL paired 20/20% 30/30% 40/40% (mmol/L) CGM-SMBG SMBG SMBG SMBG 40-400 7508 (2.2-22.2) 40-80 1699 (2.2-4.4) Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 137: System Performance In Pediatrics

    Type 1 diabetes and about 3% had Type 2 diabetes. The average age of study subjects was 12 years old, the youngest subject was 2 years of age and the oldest subject was 17 years of age. Dexcom G4 PLATINUM User’s Guide...
  • Page 138 20% of SMBG glucose values (>80 mg/dL, i.e. 4.4 mmol/L). Table 1. Percentage of CGM Readings Within SMBG Values Glucose Number of Ranges Percent Percent Percent Matchec mg/dL within 20/20 within 30/30 within 40/40 Pairs (n) (mmol/L) 40-400 1882 (2.2-22.2) Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 139 YSI values at the beginning (Day 1), middle (Day 4) and end (Day 7) of the System lifecycle. The average and median of the absolute percent differences are included in Table 5 showing consistent accuracy and sensor stability over the 7-day life of the sensor. Dexcom G4 PLATINUM User’s Guide...
  • Page 140 For a variety of reasons, the System may not display a glucose reading and readings are “skipped.” Adjusted within each system wear-day, the System provided an average of 92% of all expected glucose readings (288). Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 141: Fcc Requirements

    • Reorient or relocate the receiving antenna. Dexcom G4 PLATINUM User’s Guide...
  • Page 142 • Connect the equipment into an outlet on a circuit different from that to which the receiver is connected. • Consult the dealer or an experienced radio/TV technician for help. This portable transmitter with its antenna complies with FCC/IC RF exposure limits for general population/uncontrolled exposure. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 143: Chapter 12: Appendix

    APPENDIX Dexcom G4 PLATINUM User’s Guide...
  • Page 144 Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 145: Appendix I, Receiver Alerts, Alarm And Prompts

    Alerts. alert level. The receiver battery Vibrates once Yes, at 10% battery is low. Charge your at 20% battery battery alert receiver as soon as capacity left. capacity left. possible when you see this alert. Dexcom G4 PLATINUM User’s Guide...
  • Page 146 Wait 15 more minutes. If error screen still appears enter 1 more blood glucose value. Wait 15 minutes. If no sensor glucose readings appear on the receiver, the sensor needs to be replaced. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 147 5 minutes until calibrate. confirmed. Additional The receiver needs 1 vibrate then Yes, every startup 1 additional blood vibrate/beep every 15 minutes. calibration glucose value to 5 minutes until prompt complete startup confirmed. calibration. Dexcom G4 PLATINUM User’s Guide...
  • Page 148 (2 times) or until confirmed. Rapid fall Your glucose levels 3 vibrates then 3 alert are falling fast at vibrates/3 beeps 0.17 mmol/L per every 5 minutes minute or more. (2 times) or until confirmed. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 149 5 minutes expiration (2 times). alert Sensor The sensor is not 1 vibrate then Yes, 2 failed alert working properly. vibrate/beep re-alerts in every 5 minutes the next 10 (2 times). minutes for 30 minutes. Dexcom G4 PLATINUM User’s Guide...
  • Page 150 Replace the every 5 minutes transmitter as (2 times). soon as possible. Transmitter The transmitter 1 vibrate then failed alert has failed. vibrate/beep Replace the every 5 minutes transmitter (2 times). immediately. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 151 • 0-hour sensor expiration alert - You will see this alert in the normal course of using a sensor for seven days. Other alerts and alarms cannot be safely verified by the user. Dexcom G4 PLATINUM User’s Guide...
  • Page 152: Glossary

    The graph shows glucose trends over the amount of time shown on the screen. mmol/L Millimoles per liter. The international standard unit of measuring blood glucose levels. Profiles Sound pattern and volume level settings for your alerts. Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 153 A sensor glucose reading shown on your receiver. This reading is given in mmol/L units and is updated every 5 minutes. Transmitter The Dexcom G4 PLATINUM System component that snaps into the sensor pod and wirelessly sends glucose information to your receiver.
  • Page 154: Symbols Used In Labeling

    The following symbols may be found on the sensor, transmitter, and receiver package labels. These symbols tell you about the proper and safe use of the Dexcom G4 PLATINUM CGM System. Some of these symbols may not have meaning in your region, and are listed for informational purposes only. This table shows what each symbol means.
  • Page 155 Electrical Equipment Do Not Use if Package is Designed Primarily for Damaged Indoor Use Input Ship By Date Marking Certifies that the device meets the Keep Dry CSA-C22.2 No. IEC 60601-1(ed.3) and IEC 60601-1-11(ed. 1) Standard Dexcom G4 PLATINUM User’s Guide...
  • Page 156 THIS PAGE LEFT INTENTIONALLY BLANK Dexcom G4 PLATINUM Continuous Glucose Monitoring System...
  • Page 157 Dexcom G4 PLATINUM User’s Guide...
  • Page 158 Dexcom, Inc. MediTech Strategic Consutants B.V. 6340 Sequence Drive Maastrichterlaan 127-129 San Diego, CA 92121 USA NL - 6291 EN Vaals +1.858.200.0200 Tel. +31.43.306.3320 0086 www.dexcom.com Fax. +31.43.306.3338 Outside US: Contact your local distributor LBL-012080 Rev 04 MT22346...

Table of Contents