ANIMAS® VIBE™ SYSTEM
C H AP TE R 11 - P U M P S AF ET Y S Y ST EM A N D A L A RM S
Alerts, Warnings and Alarms
Warning: Bolus Delivery Canceled
Cause
Effect
Message
Action
Beeps/Vib
Alarm: Occlusion
Cause
Effect
Message
Action
Beeps/Vib
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 89
User pressed function button on pump during bolus
delivery.
Bolus delivery stopped.
Every 3 minutes or when awakened manually.
Press
to confirm. If button was pressed accidentally,
repeat steps (see Chapter 4 in Section I) to deliver
remaining insulin units.
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vib within one hour.
Occlusion/blockage detected in the insulin delivery
path.
All insulin deliveries stop.
Continuous until confirmed.
Press
to confirm. Disconnect and prime to clear
occlusion. Option to select "Suspend" (see Suspend
Warning screen in this chapter).
User selected, every 3 minutes until confirmed. If not
confirmed, progresses to sweep/vib within one hour.
(Once confirmed, No Prime warning triggered, see No
Prime Warning screen in this chapter.)
89
Dec 2014
Warning
Bolus
delivery
canceled by
user button
press.
Delivered:
X.XX U of X.XX U
Confirm
ALARM
OCCLUSION
DETECTED
No delivery.
Suspend
Confirm
A197 of A312
11/14/14 2:53 PM