C H AP TE R 1 4 - TR O U BL E SH O OT I NG H Y P OG L Y C EM IA , H Y P E RG L YC E MI A , P R O BL E M S W I TH
Troubleshooting Pump Operation Problems
Changing insulin delivery settings
If a basal delivery occurs while you are changing any of the pump settings that affect basal delivery
(specifically, Battery Type, IOB, Low Cartridge Warning Level, Temp Basal Alert Sound Type or any of the
delivery limit settings), but you have not confirmed the setting changes by pressing
will be used temporarily only for the basal delivery that was in process at the time the changes were
entered. If you do not confirm your setting changes by pressing
changes will be lost, and the pump will revert to the last confirmed settings for the next basal delivery.
The same situation will occur, if the pump enters into sleep mode while you are changing the pump
settings but before you confirm the changes by pressing
will not be saved, and, when you awaken the pump, the setting will revert to the last confirmed settings
saved in the pump.
Once confirmed by pressing
next basal delivery.
Remember to confirm your settings. You can also suspend the pump while you are changing settings and
then resume insulin delivery after all changes have been confirmed.
Changing Occlusion Sensitivity level
When changing the Occlusion Sensitivity Level setting on the SETUP ADV 6 screen, be sure to confirm the
change by pressing
new setting, and the change will not be stored.
If the pump enters into sleep mode before you confirm the setting changes, you should edit and
confirm the desired setting, which will take effect after your next Prime/Rewind sequence. Remember to
disconnect before priming and follow the instructions for Prime/Rewind provided in Chapter 3.
Basal History Record
If your basal rate is greater than 20 U/Hr and you receive a Call Service Alarm, this may result in a single
incorrect Basal History record. This does not impact the programmed basal rate or the calculation of the
actual amount delivered; your programmed basal rate will remain unchanged, and your Total Daily Dose
will reflect the actual amount of insulin delivered.
Contact 24-hour Customer Service for assistance if you receive a Call Service Alarm.
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 103
Y OU R I N F U SI O N SET S/ SIT ES , A N D P UM P O P E R A T IO NS
, new settings will be saved and stored in the pump and utilized for the
before the pump enters into sleep mode. Otherwise, the pump may not display the
when you enter the changes, the
. The change that went into effect temporarily
103
Dec 2014
, the new settings
A211 of A312
11/14/14 2:53 PM