ANIMAS® VIBE™ SYSTEM
C H AP TE R 3 - GE T TI N G Y O U R PU M P R EA D Y
10.
With "Go Prime" highlighted on the DELIVER PRIME screen, press
display the PRIMING ACTIVE screen and begin priming.
m
WARNING: NEVER start any part of the Prime/Rewind sequence on your pump while the
infusion set is connected to your body. Always disconnect before you begin the sequence.
The Prime/Rewind sequence includes steps for rewinding the pump motor, loading an insulin
cartridge, and priming the infusion set tubing. Failure to disconnect your infusion set from your
body before performing these steps can result in over delivery of insulin.
11.
With the PRIMING ACTIVE screen displayed,
press and hold the
of insulin come out the end of your infusion set. This
means your tubing is primed.
12.
Release the
DONE screen. After a few seconds, the ezPrime
screen will appear and "Fill Cannula" is highlighted.
MAKE SURE your infusion set is properly primed by confirming that you have seen 5 drops of insulin
come out the end of your infusion set. The amount of Prime insulin displayed on the PRIMING DONE
screen may differ from the amount displayed during the Priming procedure by ±2U. If additional
priming is required, select "Prime" a second time from the ezPrime screen and repeat steps for
priming until you are sure 5 drops of insulin come out the end of your infusion set.
The maximum Prime amount is 20U at a time. The amount of Prime insulin displayed on the PRIMING
DONE screen may differ from the amount displayed during the Priming procedure by ±2U. If additional
priming is required, select "Prime" a second time from the ezPrime screen and repeat the steps for
priming until you are sure 5 drops of insulin come out the end of your infusion set.
Refer to the Instructions for Use included with your infusion set for proper insertion guidelines. See
Selecting the Infusion Site and Inserting the Infusion Set in this chapter.
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 22
button until you see 5 drops
button to display the PRIMING
to
PRIMING
ACTIVE
Press AND HOLD
OK button
while priming
tubing.
Insulin: 182U
------
------
------
>>
>>
>>
22
Dec 2014
DELIVER
PRIME
Press AND HOLD
OK button
while priming
tubing.
Insulin: 200U
Go Prime
Cancel
PRIMING
DONE
Primed
18U
Remaining
182U
A130 of A312
11/14/14 2:52 PM