Changing the Battery

Always disconnect from infusion site prior to changing the battery.
After you replace the battery:
• A full rewind and prime sequence is required. See Priming your Pump and Infusion Set in this chapter.
• The Insulin on Board (IOB) calculation is reset to zero (0.00U). For example, if the user changes the
battery before the IOB amount decreases to 0.00U, the IOB will automatically be set to 0.00U.
The IOB calculation will reflect the correct bolus amount at the next bolus setting.
• All bolus deliveries are canceled at the time of a battery change (Audio Bolus, Combo Bolus and
Normal Bolus). Combo Bolus returns to the factory default for duration and split settings. For more
information regarding the types of bolus delivery, including Combo Bolus, refer to Chapter 10 in
Section I.
• The Combo bolus returns to the factory set default duration and split.
• You should review your basal program settings to make sure they have been saved in the pump.
• Basal program delivery will automatically resume.
• Any Temp Basal in effect before the battery change will be canceled.
• To resume your Temp Basal program, you will need to re-enter and initiate your Temp Basal settings.
• Any active CGM session in effect before the battery change will automatically resume.
m
WARNING:
• MAKE SURE you understand and are prepared for what actions you might need to take
when there is little or no battery power remaining in your pump. Make sure to have an
alternate method for administering insulin, such as pens and syringes, if you are likely to be
disconnected for an extended period of time. Contact your HCP for making insulin adjustments
with a pen or syringe when disconnected for an extended period of time.
• Low Battery Warning will appear when the remaining battery life approaches 30 minutes. Upon
receiving this warning the user should replace the battery to avoid an interruption in insulin
delivery.
• You can remove the pump battery for up to 12 hours without having to reset the time and date.
Once you re-insert the battery, after the 12-hour period, the pump will prompt you to enter the
time and date. You must reset time and date to continue using the pump.
• If you are using the Dexcom G4 PLATINUM Sensor and Transmitter with your pump, the
Replace Battery alarm will stop all CGM functions, and no further CGM readings will be
displayed until the battery is replaced. See Section II.
P130007
41028900 RevB_OB_VIBE_US_EN_MG_R4.indd 12
12
Dec 2014
A120 of A312
11/14/14 2:52 PM