Using N8480 Series Power Sensors
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9. These tables do not contain any data when the power meter is
shipped from the factory.
Editing frequency dependent offset tables requires the following steps:
1 Identify and select the table you want to edit
2 Rename the table
3 Enter the frequency and offset pairs
4 Save the table
Procedure
First, select the table you want to edit as follows:
1 Press

Figure 6-11 "Sensor Tbls" screen

2 Choose the table you want to edit using the
164
2
The 8482B and 8482H power sensors use the same data as the 8482A.
Tables
,
screen.
Edit Table
to display the Edit Cal screen as shown in
Sensor Cal Tables
,
N1913/1914A EPM Series Power Meters User's Guide
to display the Sensor Tbls
and
keys. Press
Figure 6-
12.